SERVICE AND TROUBLESHOOTING
WHY DOES THE LASER NOT FIRE WHEN I PRESS THE ACTIVATION PEDAL?
CAUSE:
1 - Activation pedal not connected.
2 - Activation pedal AA Batteries are too low to operate.
3 - Laser is in Standby mode.
SOLUTION:
1 - Check to see if the Bluetooth indicator on the activation pedal is on, and if
there is a Bluetooth indicator on the top right of the display. If either indicator is
off, re-sync the footpedal by turning the laser unit off, waiting 10 seconds, and
turning it back on. Step on the activation pedal after a wavelength mode is selected
to re-sync.
2 - Replace the two AA batteries in the activation pedal.
3 - Touch the ACTIVE/STANDBY selection to activate the laser .
I CAN HEAR THE LASER FIRING BUT IT IS CUTTING SLOWLY OR NOT CUTTING AT ALL
CAUSE:
1 - Fiber tip is not initiated.
2 - Fiber tip is not in contact with tissue.
3 - Power setting is too low.
SOLUTION:
1 - The fiber tip should be initiated for any procedures where removal of tissue is
necessary.
2 - The fiber tip should be in contact with target tissues where removal of tissue is
necessary.
3 - The laser unit should be set at the proper power setting for the procedure. The
preset procedures are a good reference point.
41