CLEANING AND STERILIZATION PROCEDURES

GUIDELINES
The Gemini 810 + 980 diode laser is not supplied in sterile condition, nor must it be sterilized before
use with the exception of the handpiece. The following cleaning and sterilization procedures are
recommended before the initial use and after each subsequent use:
1. The disposable fiber tips are supplied non-sterile by the manufacturer and should be wiped
with Isopropyl alcohol wipes by the operator prior to use. The tips are to be discarded in
an infectious waste container (SHARPS) after each use. There is no re-use or re-processing
procedure indicated for the disposable fiber tips.
2. The aluminum handpiece is also provided non-sterile by the manufacturer and should be
cleaned and sterilized prior to initial use and after each use following these instructions:
CLEANING
Warning: The Gemini 810 + 980 diode laser and its components cannot be
cleaned with an automated cleaning process.
The cleaning process is intended to remove blood, protein, and other potential contaminants from
the surfaces and crevices of reusable accessories. This process may also reduce the quantity of
particles, microorganisms, and pathogens present. Cleaning must be performed within a maximum
of 1 hour after the procedure and always prior to sterilization:
1. After use, carefully remove the disposable fiber tip from the handpiece and dispose of in an
infectious waste container (SHARPS).
2. Clean the handpiece and attached fiber cable by using one CaviWipes® towelette, or
equivalent product, to completely pre-clean exposed areas of all gross debris. Be sure to wipe
the threaded area where the disposable tip attaches.
3. Remove the handpiece shell following the instructions on the next page. Use a fresh towelette
to pre-clean the newly exposed handpiece shell thread and the end cap (starting at the o-ring
and wiping towards the fiber cable).
4. Use a new towelette to thoroughly wet all pre-cleaned areas, keeping all areas wet for 2
minutes at room temperature (68° F/20° C). Repeated use of towelettes may be required to
ensure that the surfaces remain visibly wet.
5. Visually inspect the handpiece to ensure there is no visible debris remaining. If necessary,
continue wiping with CaviWipes until all visible debris is removed.
6. Wipe all exposed areas of the handpiece with isopropyll alcohol wipes to remove any residue
left by the CaviWipes®
STEAM STERILIZATION
The steam sterilization process is intended to destroy infectious microorganisms and pathogens.
Always perform the sterilization procedure immediately after cleaning and prior to use, and only
use FDA-cleared (USA) or CE-marked (Europe) sterilization accessories such as sterilization pouches
and autoclave trays.