CLEANING AND STERILIZATION PROCEDURES
1. Place the handpiece shell in a separate single-wrap, self-seal autoclave pouch.
2. Place on an autoclave tray with paper side up; do not stack other instruments on top of the
pouch.
3. Place the tray inside the autoclave chamber and set the cycle to 135° C (275° F) for a minimum
of 15 minutes, with a dry time of 30 minutes.
4. Once the cycle is completed, remove the tray and let the sterilized item cool and dry. The
handpiece must remain in the sterilization pouch until used in order to maintain sterility.
5. Reassemble the handpiece following the instructions below.
Turn aluminum handpiece
shell counter-clockwise
to remove
To reassemble, slide the handpiece shell onto the handpiece body and turn clockwise to tighten.
NOTE: The exterior of the laser unit is not routinely contaminated by procedures. The keypad and
electroluminescent display should be covered with a protective clear adhesive barrier film, replaceable
after each patient. If the exterior of the laser unit becomes contaminated, it should be wiped down with
CaviWipes®, or equivalent product, then re-covered with a new protective plastic cover.
DO NOT spray any disinfectant directly on the laser unit, because it could damage the transparent
electroluminescent display.
DO NOT use abrasive materials to clean the laser or the display.
NEVER point the laser tip directly at the face, eyes, or skin of anyone while emitting energy.
Remove aluminum handpiece
shell for cleaning and
sterilization.
Use extreme care to avoid
accidentally damaging the
LED lens.