INSTALLATION Unpack the generator from the An earth lead is connected Using the Hex-Key shipping crate and place on a flat to the inside of the panel. provided in the surface. To remove the transit This can be removed by fittings kit, turn the brackets, firstly remove the LH side pulling it off the spade...
Page 3
Remove the RH side panel as before, Repeat steps 4 and 5 to remove the RH transit taking care to remove the earth lead bracket located under the RH compressor. located on the inside of the door panel. This will pull out in the same way. **NOTE** Do not discard the transit brackets or screws as these will be needed again if the unit is to be transported from its current location.
Page 4
If the displayed voltage is 219V or less* we would highly recommend fitting a transformer. This can be ordered directly from <= 219V Peak Scientific, Peak part number is... 06-3200 – Dual Tap Transformer 200V – 230V >= 194V IG-Genius1022 Rev.
Page 5
Switch the generator power OFF at the Remove the mains cable completely from the switch on the front panel. power supply. Re-attach both the RH & LH side panels to the generator, taking care to ensure the Once the side panels are securely earth lead’s are reconnected correctly.
Page 6
The ‘NITROGEN’ pressure gauge on the After 30 minutes of running, the generator front of the generator will gradually increase can be turned off. Allow the stored gas in the from 0 to 116psi as the pressure in the internal tanks to dissipate. internal storage tanks increases.
Page 7
CD supplied in the fittings kit. Please keep this for future reference. It is very important to register your generator with PEAK SCIENTIFIC. This will initiate your warranty entitlement. Please use the form on the next page to register your generator.