Technical Specifications; Input And Output - Siemens SITRANS P DS III Manual

Sitrans tank level transmitter
Hide thumbs Also See for SITRANS P DS III:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
6.5
Disposal
Note
Special disposal required
The device includes components that require special disposal.
● Dispose of the device properly and environmentally through a local waste disposal contractor.
7

Technical specifications

7.1

Input and output

Gauge pressure input
Measured variable
Span (continuously
adjustable) or measuring
range, max. operating
pressure (in accordance
with 97/23/EC Pressure
Equipment Directive) and
max. test pressure (in
accordance with
DIN 16086) (max. 120 bar
for oxygen measurement)
SITRANS P DS III (7MF4.33.. 7MF4.34.. 7MF4.35..)
A5E03434626-03, 06/2013
Devices identified by this symbol may not be disposed of in the municipal waste
disposal services under observance of the Directive 2002/96/EC on waste electronic
and electrical equipment (WEEE).
They can be returned to the supplier within the EC or to a locally approved disposal
service. Observe the specific regulations valid in your country.
HART
Gauge pressure
Span
Maximum
operating
pressure
MAWP (PS)
0.01 ... 1 bar g
4 bar g
(0.15 ...
(58 psi g)
14.5 psi g)
0.04 ... 4 bar g
7 bar g
(0.58 ...
(102 psi g)
58 psi g)
0.16 ...
21 bar g
16 bar g (2.3
(305 psi g)
... 232 psi g)
0.63 ... 63 bar
67 bar g
g (9.1 ...
(972 psi g)
914 psi g)
1.6 ...
167 bar g
160 bar g (23
(2422 psi g)
... 2321 psi g)
4 ... 400 bar g
400 bar g
(58 ...
(5802 psi g)
5802 psi g)
7.0 ...
800 bar g
700 bar g
(11603 psi g)
(102 ...
10153 psi g)
PROFIBUS PA and FOUNDATION
Maximum
Measuring
test
range
pressure
6 bar g
1 bar g
(87 psi g)
(14.5 psi g)
10 bar g
4 bar g
(145 psi g)
(58 psi g)
32 bar g
16 bar g
(464 psi g)
(232 psi g)
100 bar g
63 bar g
(1450 psi g)
(914 psi g)
250 bar g
160 bar g
(3626 psi g)
(2321 psi g)
600 bar g
400 bar g
(8702 psi g)
(5802 psi g)
800 bar g
700 bar g
(11603 psi g
(10153 psi g)
)
Fieldbus
TM
Maximum
Maximum
operating
test pressure
pressure
4 bar g
6 bar g
(58 psi g)
(87 psi g)
7 bar g
10 bar g
(102 psi g)
(145 psi g)
21 bar g
32 bar g
(305 psi g)
(464 psi g)
67 bar g
100 bar g
(972 psi g)
(1450 psi g)
167 bar g
250 bar g
(2422 psi g)
(3626 psi g)
400 bar g
600 bar g
(5802 psi g)
(8702 psi g)
800 bar g
800 bar g
(11603 psi g)
(11603 psi g)
35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mf4.34 series7mf4.33 series7mf4.35 series

Table of Contents