Read
Source
Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude is
adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight negative
offset of approximately -0.1V is present to assure zero crossing.
Ohms low
Ohms high
Ohms low
Ohms high
Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.
Read (mV)
Source (mV)
Maximum current output in voltage ranges is 1mA with an output impedance of <= 1Ω
TC Type
J
K
50
Find Quality Products Online at:
Table 12: Frequency Measurement/Source
2.0CPM - 600.0CPM
1.00KHz - 10.00KHz
2.0CPM - 600.0CPM
1.00KHz - 10.00KHz
Table 13: Resistance Measurement
Table 14: Resistance Source
Range
5.0Ω - 400.0Ω
5.0Ω - 400.0Ω
400Ω - 1500Ω
1500Ω - 4000Ω
Table 15: Thermocouple Measurement/Source
-10.000mV - 75.000mV
-10.000mV - 75.000mV
Table 16: Thermocouple Read and Source (errors in °C)
Range (°C)
Minimum
-210.0
-150.0
-200.0
-100.0
600.0
1000.0
GlobalTestSupply
www.
Range
1.0Hz - 1000.0Hz
1.0Hz - 1000.0Hz
Range
0.00Ω - 400.0Ω
401.0Ω - 4000.0Ω
Excitation Current
0.1mA - 0.5mA
0.5mA - 3mA
0.05mA - 0.8mA
0.05mA - 0.4mA
Range
Maximum
-150.0
1200.0
-100.0
600.0
1000.0
1372.0
Accuracy (% of reading ± floor)
0.05% ± 0.1CPM
0.05% ± 0.1Hz
0.05% ± 0.01KHz
0.05%
0.05%
0.250%
Accuracy (% of reading ± floor)
0.025% ± 0.05Ω
0.025% ± 0.5Ω
Accuracy (% of reading ± floor)
0.025% ± 0.1Ω
0.025% ± 0.05Ω
0.025% ± 0.5Ω
0.025% ± 0.5Ω
Accuracy (% of reading ± floor)
0.02% ± 10μV
0.02% ± 10μV
Accuracy (°C)
CJC OFF
CJC ON
0.4
0.6
0.2
0.4
0.5
0.7
0.2
0.4
0.3
0.5
0.4
0.6