Martel MC1210 Reference Manual page 30

Table of Contents

Advertisement

7. At the top of the display it will now read, "SW OPENED AT" and give you the pressure that
the switch opened at.
8. Press the "NEXT" option to view when the switch closed, and the dead band.
Important NOTE: The previous example uses a normally closed switch. The basic
procedure is still the same for a normally open switch, the display will just read "OPEN"
instead of "CLOSE".
30
Find Quality Products Online at:
SW TEST OPEN
------
28.000
LIGHT
MENU
SW OPENED AT
30.000
28.000
LIGHT
MENU
GlobalTestSupply
www.
psi
psi
psi
psi
.com
sales@GlobalTestSupply.com

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents