Center Frequency - HP 85662A Troubleshooting And Repair Manual

Spectrum analyzer if-display section
Hide thumbs Also See for 85662A:
Table of Contents

Advertisement

A4A5 Step Gain, Troubleshooting
Operation of the Step Gains and AO.O to A15.9 Attenuator can be determined from the
display by several easy tests.
The Step Gains are checked in the following way. Connect a 10 dB step attenuator between
CAL OUTPUT and the input to the analyzer. Set this attenuator to 90 dB, press Instrument
Preset, and key in the following:
[CENTER
FREQUENCY!
equal to the CAL OUTPUT frequency
f FREQUENCY SPAN)
0 Hz
[RES BW)
[ATTEN)
[VIDEO B W )
( S W E E P )
[REFERENCE LEVEL)
-10
dBm
1kHz
0dB
10 Hz
20 s
Press the
[SINGLE)
sweep key and press the DATA STEP QQ key to display each step. Figure 5
should be displayed. If one or more of the steps is missing, refer to Note 8 for the Step Gain
truth table to determine which stage is faulty. 0 V in the table indicates that a control line is
active.
1 0
d B /
R E F - 9 9 . 9 dBm
ATTEN 0 dB
- 1 0
dBm
AND SG20-1
CENTER 2 0 . 0 0 0 0 0 0 MHz
RES BW 1 kHz
VBW 10 Hz
SPAN 0 Hz
S W P 2 0
sec
NOTE: C e n t e r
F r e q u e n c y r e a d o u t w i l l be 1 0 0 . 0 0 0 000 MHz for an HP8566B
Figure 5. 10 dB Step Gain Test
If a component has been replaced, a simple calibration/diagnostic program exists. The
calibration routine is started by connecting a cable from CAL OUTPUT to the input of the
analyzer and then pressing
(SHIFT") FRIQUENCY SPAN)
When this program has finished, press
fhiiiFT)
[UNE")
TRIGGER w ichwill list the calibration data. (Refer to Figure 6). Lines 19
A4A5
5