Commissioning
 ORP electrodeSensoLyt
 Operating manual
Prepare the electrode for measuring as follows:
 Remove the watering cap from the electrode tip. Possible salt deposits in
the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.
Please keep the watering cap. It is required for the electrode
to be stored. Always keep the watering cap clean.
 Connect the sensor to the meter.
The sensor is immediately ready to measure.
Types of connection:
– via a connecting cable (accessory) to a free IDS connector on the meter
or
– wireless via an IDS WLM-S adapter (accessory) to a WLM-capable
meter
or
– via an IDS sensor slot of a multi parameter probe
(connected to a meter)
Accessories for the connection of the SensoLyt
the meter are listed in chapter 7 W
To open and close the IDS plug-in position please note the section 2.1
O
PENING AND CLOSING THE PLUG CONNECTION
2.3
Calibrating, measuring
Remove the watering cap for calibrating and measuring. More
detailed information is given in the operating manual of the meter.
U
= U
+ U
H
Meas
Ref
with:
U
= ORP, referring to the normal hydrogen electrode
H
U
= Measured ORP
Meas
U
= Voltage of the reference system compared to the normal
Ref
hydrogen electrode
U
depends on the reference system and temperature and can be taken
Ref
from the following table (see also DIN 38404-6 for the system Ag/AgCl/
saturated KCl):
®