Operating & Maintenance Tips; Appendix-Non-Conventional Support Rings - General EMHT40 User Manual

Hardness tester astm rated pen style
Table of Contents

Advertisement

Auto Power Off Trigger: 2 minutes of control button inactivity
Power Source: 3.7V rechargeable Li-ion-battery
Dimensions: 6.2 x 1.6 x 1 in. (158 x 41 x 26mm)
Weight: 4.2 oz. (120g)
OPERATING & MAINTENANCE TIPS
The EMHT40's battery should be kept charged and charged frequently to avoid damaging it.
Using the AC adaptor, it takes about 4 hours to fully charge the battery from a completely
discharged state. Trickle-charging the battery from a PC's USB port takes longer.
Over time, the tip of the impact body will abrade and reduce the EMHT40's measurement
accuracy. Whenever measured values seem unusually large, calibrate the instrument by
following the procedure described on pages 16 and 17.
APPENDIX: Non-conventional support rings
No. Code
Type
Sketch
Remarks
-21-