Page 1
Operating Instructions Aqua-Purificator G 7895 and conductance measuring module C To prevent accidents and machine damage read these instructions before installation or use. M.-Nr. 06 159 860...
Guide to the machine a Machine lid d Power cord b Door handle e "Entrance" hose (blue) connects to tap water c Conductivity display meter f Hose "H O pur" (green) connects to labwasher...
Page 4
Guide to the machine g Water demineralization cartridge h Connection "Exit" i Connection "Entrance"...
IMPORTANT SAFETY INSTRUCTIONS Please pay attention to the following This machine conforms to current notes, to maintain safe procedures. safety requirements. However, The machine should be inappropriate use can lead to commissioned and then personal injury and property maintained only by a Miele Authorized damage.
Page 6
IMPORTANT SAFETY INSTRUCTIONS Never clean the machine near or Using accessories with a water or high pressure hose. Only specific equipment made by Before servicing, disconnect the Miele should be connected to this power supply by either removing machine. Consult Miele on the type and the fuse, unplugging the unit or application of such equipment.
Description of the machine The Aqua Purificator G 7895 is To ease the emptying and refilling of the operated using the E 318 water water demineralization cartridge E 318, demineralization cartridge: we recommend as additional equipment to the cartridge E 318, the The cartridge is empty when delivered "Original Miele resin replace set E 316"...
Filling Cartridge E 318 with one-time resins E 315 Opening the lid Filling the resins into the cartridge ^ Insert the funnel into the cartridge opening. ^ Hold the lid, to keep it from falling into the cartridge and unscrew the winged nut.
Connection of E 318 cartridge ^ Open the machine front. ^ Unscrew the caps from the cartridge connection. ^ Screw the hoses onto: "Entrance" (blue) to "Entrance", "H O pur" (green) to "Exit". ^ See graphic for hose location.
Connection ^ Connect the "H O pur" hose (green) Priming the cartridge to the DI water intake hose of the respective machine. ^ Connect the "Entrance" hose (blue) with the on-site cold water valve. See the Installation manual Important The water pressure (flow rate) must be between 37 and 145 kPa.
Conductance display The conductance display shows the Uninstall the cartridge E 318 conductance of the demineralized ^ Turn off the water valves. water, measured in ZS/cm (micro ^ If available, open the water valve for siemens per centimeter) and is a manual DI water extraction.
Conductance measure module The conductance measure module has the same function as the built-in conductance measure module in the Aqua Purificator G 7895. The water conductance is measured and a quality control of the demineralized water enabled. ^ The connection of the conductance...
Electrical connection All electrical work must be carried out by a suitably qualified electrician in accordance with local and national safety regulations. ^ The unit is supplied without a plug. A proper plug for the receptacle available is user supplied. ^ The electrical requirement is 220V.
After sales service Should you have any difficulties please contact your authorized Service Dealer or the Miele Service Department. See back cover for addresses. Please quote the model and serial number of your machine. This information can be found on the data plate (see also “Electrical connection”).
Caring for the environment Disposal of the packing Disposal of the appliance material Appliances contain materials which can be recycled. Please contact your local The protective packing materials are authorities about recycling in your area. environmentally safe and can be recycled.
Page 16
All rights reserved / 1104 M.-Nr. 06 159 860 / 00 This paper has been bleached without the use of chlorine.
Need help?
Do you have a question about the G 7895 and is the answer not in the manual?
Questions and answers