Reassembling The Mask; Technical Specifications; Dead Space Information - ResMed Ultra Mirage II User Manual

Nasal mask
Hide thumbs Also See for Ultra Mirage II:
Table of Contents

Advertisement

If any visible deterioration of a mask
component is apparent (cracking,
crazing, tears etc), the mask
component should be discarded and
replaced.
R
EPROCESSING THE
P
BETWEEN
ATIENTS
The mask should be reprocessed when
used between patients.
Cleaning, disinfection and sterilization
instructions are available from the ResMed
website, www resmed.com/masks/
sterilization/americas.
If you do not have Internet access, please
contact your ResMed representative.
R
EASSEMBLING THE
See section E on the illustrations sheet.
1 Insert the ports cap into the ports. Make
sure the tabs of the ports cap face away
from the mask.
2 Insert the mask elbow into the mask
frame and secure it with the elbow clip
on the inside of the frame (Fig. E-1). The
flat side of the elbow clip must face the
mask frame. If you have fitted the mask
elbow correctly, it should not move
when you try to firmly pull it away from
the frame.
3 Grasp the edges of the vent cover on
both sides of the "Click" sticker at the
same time. Firmly push the vent cover
onto the mask elbow to make sure it is
clicked securely in place (Fig. E-2).
4 Push the swivel onto the end of the
mask elbow.
5 Attach the forehead pads to the
forehead support by inserting the pad
plugs into the two holes on the back of
the forehead support (Fig E-3). Push the
pads firmly against the forehead support
M
ASK
M
ASK
to ensure that the pads fit securely into
the forehead support.
6 Reconnect the forehead support to the
mask frame. Depress the tab and move
the forehead support into one of the
four tab positions (Fig. E-4).
7 Fit the grooved edge of the nasal
cushion to the mask frame. Fit the top
section first, then work your way around
the mask, pushing the cushion onto the
frame, ensuring a good seal has been
achieved. Push the cushion clip over the
cushion onto the mask frame, ensuring it
clicks into place in all locations (Fig. E-5).
8 Connect the headgear clips to the mask
frame. Thread the headgear straps
through the slots on the forehead
support and the headgear clips. The
U
M
headgear label should
LTRA
IRAGE
face outwards and be near your neck
when fitted. Fasten the Velcro straps.
T
S
ECHNICAL
PECIFICATIONS
P
–F
RESSURE
LOW
See section A on the illustrations sheet.
D
S
I
EAD
PACE
NFORMATION
Dead space is the empty volume of the
mask up to the swivel.
The dead space of the mask varies
according to cushion sizes but is less than
135 mL.
Note:
The manufacturer reserves the right
to change these specifications without notice.