Flexim FLUXUS F704**-NN Series User Manual

Flexim FLUXUS F704**-NN Series User Manual

Ultrasonic flowmeter for liquids
Table of Contents

Advertisement

Ultrasonic Flowmeter
for Liquids
FLUXUS F704**-NN
FLUXUS F704**-A2
FLUXUS F704**-F2
FLUXUS F706**-NN
FLUXUS F706**-A2
FLUXUS F706**-F2
UMFLUXUS_F7V4-6-2EN
FLUXUS F705**-NN
FLUXUS F705**-A2
FLUXUS F705**-F2
FLUXUS F709**-NN
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Flexim FLUXUS F704**-NN Series

  • Page 1 User Manual UMFLUXUS_F7V4-6-2EN Ultrasonic Flowmeter for Liquids FLUXUS F704**-NN FLUXUS F705**-NN FLUXUS F704**-A2 FLUXUS F705**-A2 FLUXUS F704**-F2 FLUXUS F705**-F2 FLUXUS F706**-NN FLUXUS F709**-NN FLUXUS F706**-A2 FLUXUS F706**-F2...
  • Page 2 FLUXUS is a registered trademark of FLEXIM GmbH. FLEXIM GmbH Wolfener Straße 36 12681 Berlin Germany Phone: +49 (30) 936 67 660 Fax: +49 (30) 936 67 680 E-mail: flexim@flexim.de www.flexim.com User manual for FLUXUS F70x UMFLUXUS_F7V4-6-2EN, 2017-10-01 Copyright (©) FLEXIM GmbH 2017...
  • Page 3 FLUXUS F70x Die Sprache, in der die Anzeigen auf dem Messumformer erscheinen, kann eingestellt werden (siehe Abschnitt 8.5). The transmitter can be operated in the language of your choice (see section 8.5). Il est possible de sélectionner la langue utilisée par le transmetteur à l'écran (voir sec- tion 8.5).
  • Page 4 FLUXUS F70x UMFLUXUS_F7V4-6-2EN, 2017-10-01...
  • Page 5: Table Of Contents

    Table of Contents FLUXUS F70x Table of Contents Introduction ................9 Regarding this User Manual .
  • Page 6 FLUXUS F70x Table of Contents Basic Measurement ..............91 Input of the Pipe Parameters .
  • Page 7 Table of Contents FLUXUS F70x SuperUser Mode ............... 134 15.1 Activation/Deactivation .
  • Page 8 FLUXUS F70x Table of Contents UMFLUXUS_F7V4-6-2EN, 2017-10-01...
  • Page 9: Introduction

    • unsuitable or insufficient maintenance • repair of FLUXUS by unauthorized personnel FLEXIM assumes no responsibility for injury to the customer or third persons proximately caused by the material owing to defects in the product which were not predictable or for any indirect damages.
  • Page 10: Handling

    FLUXUS F70x 2 Handling Handling First Inspection The measuring instrument has already been tested thoroughly at the factory. At delivery, proceed to a visual control to make sure that no damage has occurred during transportation. Check that the specifications of the measuring instrument delivered correspond to the specifications given on the pur- chase order.
  • Page 11: General Principles

    3 General Principles FLUXUS F70x General Principles For the ultrasonic measurement of the flow rate, the flow velocity of the medium flowing in a pipe is determined. Further physical quantities (e.g., volumetric flow rate, mass flow rate, heat flow rate) are derived from the flow velocity and from additional physical quantities, if necessary.
  • Page 12 FLUXUS F70x 3 General Principles Sound speed c Speed of the propagating sound. The sound speed depends on the mechanical properties of the medium or the pipe ma- terial. In pipe materials and other solid materials, a distinction is made between the longitudinal and transversal sound speed.
  • Page 13 3 General Principles FLUXUS F70x transducer (emitter) transducer (receiver) reduction in distance in the transducer α α α pipe wall β β γ flow direction γ of the medium sound path without flow sound path with flow Fig. 3.2: Sound path of the signal in the flow direction transducer (receiver) transducer (emitter) increase in distance...
  • Page 14 FLUXUS F70x 3 General Principles 3.2.3 Measurement of the Flow Velocity in the NoiseTrek Mode When media with a high proportion of gas bubbles or solid particles are measured, the attenuation of the ultrasonic signal increases and can inhibit the propagation of the signal in the medium. A measurement in the TransitTime mode is no lon- ger possible.
  • Page 15: Measurement Arrangements

    3 General Principles FLUXUS F70x Measurement Arrangements 3.3.1 Terms and Definitions Diagonal arrangement The transducers are mounted on the opposite sides of the pipe (see Fig. 3.6). Reflection arrangement The transducers are mounted on the same side of the pipe (see Fig. 3.7). Fig.
  • Page 16 FLUXUS F70x 3 General Principles Sound beam plane The plane containing one, two or more sound paths or beams (see Fig. 3.10). 2 sound paths 2 beams in 1 plane in 1 plane Fig. 3.10: Sound paths and beams in one plane 3.3.2 Examples Diagonal arrangement with 1 beam...
  • Page 17 3 General Principles FLUXUS F70x Diagonal arrangement with 4 beams and 2 planes Reflection arrangement with 4 beams and 4 planes 4 transducer pairs 4 transducer pairs 4 sound path 8 sound paths 4 beams 4 beams 2 planes 4 planes X arrangement UMFLUXUS_F7V4-6-2EN, 2017-10-01...
  • Page 18: Description Of The Transmitter

    FLUXUS F70x 4 Description of the Transmitter Description of the Transmitter Attention! The degree of protection of the transmitter will only be ensured if the cable glands are firmly tightened and the housing is tightly screwed. Design of FLUXUS F704 The front plate has to be removed to access the command panel.
  • Page 19: Design Of Fluxus F706

    4 Description of the Transmitter FLUXUS F70x Design of FLUXUS F706 The front plate has to be removed to access the command panel. 2x 16-digit LCD display, keyboard backlit transducers serial interface RS232 channel D channel C channel B channel A power supply AGN ARS AR BGN BRS BR...
  • Page 20: Keyboard

    FLUXUS F70x 4 Description of the Transmitter Keyboard 4 Description of the Transmitter The keyboard consists of three function keys ENTER, BRK and C and ten numerical keys. Several keys have double functions. They can be used for entering data and for navigating through scroll lists. The arrow-shaped keys are used as cursor keys in the selection mode and for entering digits and letters in the input mode.
  • Page 21: Selection Of The Measuring Point

    5 Selection of the Measuring Point FLUXUS F70x Selection of the Measuring Point Attention! Observe the Safety Instructions for the Use in Explosive Atmosphere (see document SIFLUXUS). The correct selection of the measuring point is crucial for achieving reliable measurement results and a high measurement accuracy.
  • Page 22 FLUXUS F70x 5 Selection of the Measuring Point Tab. 5.1: Recommended transducer position Horizontal pipe Select a measuring point where the transducers can be mounted on the side of the pipe, allowing the sound waves to propagate in the pipe horizontally. Thus, solid at the bottom or gas bubbles at the top of the pipe will not influence the propagation of the signal.
  • Page 23: Undisturbed Flow Profile

    5 Selection of the Measuring Point FLUXUS F70x Undisturbed Flow Profile Some flow elements (elbows, slide valves, valves, control valves, pumps, reducers, diffusers, etc.) distort the flow profile in their vicinity. The axisymmetrical flow profile needed for correct measurement is no longer given. A careful selection of the measuring point helps to reduce the impact of disturbance sources.
  • Page 24 FLUXUS F70x 5 Selection of the Measuring Point Tab. 5.2: Recommended distance from disturbance sources; D – nominal pipe diameter at the measuring point, l – recommended distance between disturbance source and transducer position disturbance source: reducer inlet: l ≥ 10 D outlet: l ≥...
  • Page 25: Selection Of The Measurement Arrangement Taking Onto Account The Measuring Range And The Measuring Conditions

    5 Selection of the Measuring Point FLUXUS F70x Selection of the Measurement Arrangement Taking onto Account the Measuring Range and the Measuring Conditions Diagonal arrangement with 1 beam Reflection arrangement with 1 beam • wider flow velocity and sound speed range compared to •...
  • Page 26 FLUXUS F70x 5 Selection of the Measuring Point Diagonal arrangement with 4 beams and 2 planes Reflection arrangement with 4 beams and 4 planes • same features as diagonal arrangement with 2 beams • same features as reflection arrangement with 2 beams and 2 planes •...
  • Page 27: Selection Of The Sound Beam Plane Near An Elbow

    5 Selection of the Measuring Point FLUXUS F70x Selection of the Sound Beam Plane Near an Elbow On vertical pipes On horizontal pipes flow direction elbow plane elbow plane flow direction • The sound beam plane (see section 3.3.1) has an an- •...
  • Page 28: Installation

    FLUXUS F70x 6 Installation Installation Attention! Observe the Safety Instructions for the Use in Explosive Atmosphere (see document SIFLUXUS). Location • Select the measuring point according to the recommendations in chapter 3 and 5. • Select the location of the transmitter within cable reach of the measuring point. The ambient temperature at the location has to be within the operating temperature range of the transmitter and the trans- ducers (see Technical Specifications).
  • Page 29 6 Installation FLUXUS F70x FLUXUS F705, F706 • Fix the transmitter to the wall (see Fig. 6.2 and Fig. 6.3). Fig. 6.2: FLUXUS F705 (dimensions in mm) Fig. 6.3: FLUXUS F706 (dimensions in mm) 6.2.3 Pipe Installation Installation on a 2" pipe The mounting kit is fixed to the pipe using a shackle.
  • Page 30: Transmitter Fluxus F709

    FLUXUS F70x 6 Installation Fig. 6.5: Installation of the transmitter Fig. 6.6: Installation of the transmitter FLUXUS F705, 706 FLUXUS F704 (example with F705) Fig. 6.7: Pipe installation using tension straps Transmitter FLUXUS F709 The transmitter is designed as a 19 " module (42 HP, 3 U). Fig.
  • Page 31: Transducers

    6 Installation FLUXUS F70x Transducers Attention! Observe the Safety Instructions for the Use in Explosive Atmosphere (see document SIFLUXUS). 6.4.1 Preparation of the Pipe • The pipe has to be stable. It has to be able to withstand the pressure exerted by the transducer mounting fixture. Rust, paint or other deposits on the pipe absorb the sound signal.
  • Page 32 FLUXUS F70x 6 Installation Overview of the mounting steps • step 1 disassemble the transducer mounting fixture Variofix L • step 2 fix the clasps to the tension straps • step 3 fix one tension strap to the pipe • step 4 use screws to fix the rail to the tension strap and fix it with the second tension strap •...
  • Page 33 6 Installation FLUXUS F70x 6.4.3.1 Disassembly of Variofix L • Disassemble the transducer mounting fixture Variofix L (see Fig. 6.13). cover screw rail tension strap clamp Fig. 6.13: Disassembly of Variofix L 6.4.3.2 Fixing the Clasps to the Tension Straps Select the installation instructions that correspond to the supplied clasp: Band clamp clasp The clasp is fixed to the tension strap (see Fig.
  • Page 34 FLUXUS F70x 6 Installation 6.4.3.3 Fixing the Tension Strap to the Pipe One tension strap is fixed to the pipe (see Fig. 6.17). The second tension strap is mounted later. metal spring tension strap clamp clasp Fig. 6.17: Tension strap with tension strap clamp and metal spring on the pipe Select the installation instructions that correspond to the supplied clasp: Band clamp clasp •...
  • Page 35 6 Installation FLUXUS F70x Ratchet clasp • Insert the tension strap into the tension strap clamp and the metal spring (see Fig. 6.22). It is not necessary to mount the metal spring: – on steel pipes – on pipes with an outer pipe diameter < 80 mm or –...
  • Page 36 FLUXUS F70x 6 Installation 6.4.3.4 Fixing the Rail to the Pipe • Place one tension strap clamp in the rail (see tension strap clamp 1 in Fig. 6.25). Observe the orientation of the tension strap clamp. • Tighten the nut of tension strap clamp 1 slightly. •...
  • Page 37 6 Installation FLUXUS F70x Quick release clasp • Insert the tension strap into tension strap clamp 1 and the metal spring (see Fig. 6.29 and Fig. 6.27). • Place the tension strap around the pipe and insert it into the clasp. •...
  • Page 38 FLUXUS F70x 6 Installation screw of the clasp lever sense of rotation Fig. 6.32: Ratchet clasp with tension strap Note! To release the screw and the tension strap, press the lever down (see Fig. 6.32). • Repeat the steps for fixing the second rail (see Fig. 6.33). Fig.
  • Page 39 6 Installation FLUXUS F70x strain cover relief clamp transducer distance Fig. 6.35: Adjustment of the transducer distance • Put the covers with the transducers on the rail. • Correct the transducer distance, if necessary (see section 9.6.1 and 9.6.2). Note! Make sure that the coupling foil remains on the contact surface of the transducers.
  • Page 40 FLUXUS F70x 6 Installation Fig. 6.38: Transducer mounting fixture Variofix C (diagonal arrangement) Overview of the mounting steps • step 1 disassemble the transducer mounting fixture Variofix C • step 2 mount the tension straps (with or without clasp) and fix the rail to the tension straps with screws •...
  • Page 41 6 Installation FLUXUS F70x cover tensioning screw spring clip indentation rail spacing element tension strap clamp Fig. 6.41: Disassembly of Variofix C 6.4.4.2 Installation of the Rail Select the installation instructions that correspond to the supplied clasp: • see section Installation of the rail without a clasp •...
  • Page 42 FLUXUS F70x 6 Installation tension metal spring strap s = length of the rail - 33 mm Fig. 6.44: Tension strap with the metal spring and the tension strap clamp on the pipe • Insert the long end of the tension strap into the second slot of the tension strap clamp (see Fig. 6.44 a). •...
  • Page 43 6 Installation FLUXUS F70x • Bend the tension strap. • Insert the tension strap into part (1) of the ratchet clasp (see Fig. 6.46 b). • Tighten the tension strap. • Insert the long end of the tension strap into the tension strap clamp and the metal spring (see Fig. 6.47). It is not neces- sary to mount the metal spring: –...
  • Page 44 FLUXUS F70x 6 Installation screw of the clasp lever sense of rotation Fig. 6.50: Ratchet clasp with tension strap Note! To release the screw and the tension strap press the lever down (see Fig. 6.50). • Put the rail on the tension strap clamps (see Fig. 6.51). •...
  • Page 45 6 Installation FLUXUS F70x spacing element transducer distance engravings on hole the transducers Fig. 6.52: Transducers in the rail (spring clip not shown) cable tie spring clip tensioning screw Fig. 6.53: Transducers in the rail screw cover Fig. 6.54: Variofix C with transducer on the pipe The cover can be removed from the mounted transducer mounting fixture as follows: •...
  • Page 46: Temperature Probe

    FLUXUS F70x 6 Installation 6.4.5 Mounting with Mounting Shoe and Band Clamp Clasp • Insert the tension strap into the groove on the upper side of the mounting shoe (see Fig. 6.56). • Position the mounting shoe and the clasp on the pipe. On horizontal pipes, mount the mounting shoe on the side of the pipe, if possible.
  • Page 47 Select the installation instructions of the supplied clasp: • for the installation with a clasp see section 6.5.2.1 • for the installation with a FLEXIM clasp see section 6.5.2.2 • for the installation with a quick release clasp see section 6.5.2.3 6.5.2.1...
  • Page 48 FLUXUS F70x 6 Installation 6.5.2.3 Installation with a Quick Release Clasp • Cut the tension strap to length (pipe circumference + at least 120 mm). • Position the temperature probe on the pipe (see Fig. 6.58). • Place the tension strap around the temperature probe and the pipe. •...
  • Page 49: Connection

    7 Connection FLUXUS F70x Connection Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). For the connections of the transmitter FLUXUS F704 see section 4.1, F705 see section 4.2, F706 see section 4.3 and F709 see section 4.4. Terminal designation AGN ARS AR BGN BRS BR...
  • Page 50 FLUXUS F70x 7 Connection X_AV AGN ARS BGN BRS BR X_BV T1A T1B S1 T2A T3A T3B S3 T4A A+ B- P1+ P2+ P4+ P5a P6a P7a SA1 SA2 SA3 SA4 SB1 SB2 SB3 SB4 T1a T1b S2 T2a T3a T3b S4 T4a 101 103 P1- P2- P4- P5b P6b P7b X_AR...
  • Page 51 7 Connection FLUXUS F70x AGN ARS AR BGN BRS BR CGN CRS CR DGN DRS DR N(-) L(+) X_AV X_AR X_BV X_BR X_CV X_CR X_DV X_DR AGN ARS AR BGN BRS BR CGN CRS CR DGN DRS DR X_AV X_AR X_BV X_BR X_CV...
  • Page 52 FLUXUS F70x 7 Connection Fig. 7.4: Transmitter FLUXUS F709 UMFLUXUS_F7V4-6-2EN, 2017-10-01...
  • Page 53: Transducers

    7 Connection FLUXUS F70x Transducers Note! If transducers are replaced or added, the sensor module also has to be replaced or added (see sec- tion 7.7). It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to avoid load on the connectors.
  • Page 54 FLUXUS F70x 7 Connection AGN ARS AR BGN BRS BR T1A T1B S1 T2A T3A T3B S3 T4A A+ B- P1+ P2+ P4+ P5a P6a P7a SA1 SA2 SA3 SA4 SB1 SB2 SB3 SB4 T1a T1b S2 T2a T3a T3b S4 T4a 101 103 P1- P2- P4- P5b P6b P7b equipotential...
  • Page 55 7 Connection FLUXUS F70x 7.1.1.2 Transducer Cable with Plastic Cable Jacket and Stripped Cable Ends Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). • Remove the left blind plug for the connection of the transducer cable (see Fig. 7.7 or Fig. 7.8). •...
  • Page 56 FLUXUS F70x 7 Connection compression part basic part cap nut N E X T D I S P U L T R A S O N I C F L O W M E T E R M U X D I S P O F F B R K...
  • Page 57 7 Connection FLUXUS F70x 7.1.1.3 Transducer Cable with Stainless Steel Conduit and Stripped Cable Ends Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). • Remove the left blind plug for the connection of the transducer cable (see Fig. 7.9 or Fig. 7.10). FLUXUS F704 •...
  • Page 58 FLUXUS F70x 7 Connection AGN ARS AR BGN BRS BR T1A T1B S1 T2A T3A T3B S3 T4A A+ B- P1+ P2+ P4+ P5a P6a P7a T1a T1b S2 T2a T3a T3b S4 T4a SA1 SA2 SA3 SA4 SB1 SB2 SB3 SB4 101 103 P1- P2- P4- P5b P6b P7b N(-)
  • Page 59 7 Connection FLUXUS F70x AGN ARS AR BGN BRS BR T1A T1B S1 T2A T3A T3B S3 T4A A+ B- P1+ P2+ P4+ P5a P6a P7a SA1 SA2 SA3 SA4 SB1 SB2 SB3 SB4 T1a T1b S2 T2a T3a T3b S4 T4a 101 103 P1- P2- P4- P5b P6b P7b PE N(-) L(+)
  • Page 60 FLUXUS F70x 7 Connection FLUXUS F705 • Remove the left blind plug for the connection of extension cable (see Fig. 7.12). • Push the extension cable through the sealing ring (only for cable gland M20, not for cable gland 1/2 NPS). •...
  • Page 61 7 Connection FLUXUS F70x Tab. 7.5: Terminal assignment terminal connection measuring measuring channel A channel B X6AV X8BV SMB connector (white or marked cable) X6AR X8BR SMB connector (brown cable) AMP-Quick connector (connection of the sensor module) 7.1.2.2 Extension Cable with Plastic Cable Jacket and Stripped Cable Ends Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere"...
  • Page 62 FLUXUS F70x 7 Connection cap nut compression part basic part N E X T D I S P U L T R A S O N I C F L O W M E T E R M U X D I S P O F F B R K...
  • Page 63 7 Connection FLUXUS F70x FLUXUS F709 • Prepare the extension cable. Cut the external shield and brush it back (see Fig. 7.16). • Push the extension cable through the shield terminal to terminal strip KL6 for measuring channel A and to terminal strip KL8 for measuring channel B.
  • Page 64 FLUXUS F70x 7 Connection 7.1.3 Connection of the Transducer Cable to the Junction Box Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). For the connection of the transducer cable with SMB connectors (connection system TS) see section 7.1.3.1. For the connection of the transducer cable with a plastic cable jacket and stripped cable ends (connection system TS) see section 7.1.3.2.
  • Page 65 7 Connection FLUXUS F70x • Fix the cable gland by screwing the cap nut onto the basic part. • Connect the transducer cable to the terminals of the junction box (see Fig. 7.20 or Fig. 7.21 and Tab. 7.9). Tab. 7.9: Terminal assignment (KL1) terminal (JBP2, JBP3) terminal (JB01) connection...
  • Page 66 FLUXUS F70x 7 Connection 7.1.3.3 Transducer Cable with Stainless Steel Conduit and Stripped Cable Ends Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). • Remove the right blind plug for the connection of the transducer cable (see Fig. 7.22). •...
  • Page 67 7 Connection FLUXUS F70x 7.1.4 Connection of the Extension Cable to the Junction Box Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). • Remove the left blind plug for the connection of extension cable (see Fig. 7.23). •...
  • Page 68: Temperature Probe

    FLUXUS F70x 7 Connection Tab. 7.11: Terminal assignment (extension cable, KL2) terminal connection white or marked cable (core) white or marked cable (internal shield) brown cable (internal shield) brown cable (core) extension cable transducer cable Fig. 7.24: Connection of the extension cable and the transducer cable to the junction box JBxx (example with JB01) Temperature Probe Attention!
  • Page 69 7 Connection FLUXUS F70x Tab. 7.12: Connection systems direct connection connection with connection with extension cable junction box transmitter transmitter transmitter transmitter FLUXUS F709: Connection to the transmitter without cable gland For the assignment and the activation of the temperature inputs see chapter 17. 7.2.1 Direct Connection of the Temperature Probe Attention!
  • Page 70 FLUXUS F70x 7 Connection FLUXUS F705, F706 • Remove the second blind plug on the right for the connection of the temperature probe (see Fig. 7.27). • Open the cable gland of the temperature probe. The compression part remains in the cap nut (see Fig. 7.28 and Tab. 7.12).
  • Page 71 7 Connection FLUXUS F70x FLUXUS F709 • Prepare the cable of the temperature probe (see Fig. 7.31). • Connect the temperature probe to the terminals of the transmitter (see Fig. 7.4, Fig. 7.30 and Tab. 7.14). temperature probe C H A N N E L K L 1 K L 3 K L 5...
  • Page 72 FLUXUS F70x 7 Connection FLUXUS F705, F706 • Remove the second blind plug on the right for the connection of the extension cable (see Fig. 7.27). • Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.28 and Tab. 7.12). •...
  • Page 73 7 Connection FLUXUS F70x FLUXUS F705, F706 • Remove the second blind plug on the right for the connection of the extension cable (see Fig. 7.27). • Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.28 and Tab. 7.12). •...
  • Page 74 FLUXUS F70x 7 Connection temperature probe extension cable Fig. 7.34: Junction box JBT2, JBT3 Temperature probe • Remove the blind plug for the connection of the temperature probe (see Fig. 7.34). • Open the cable gland of the temperature probe (see Fig. 7.35 and Tab. 7.12). The compression part remains in the cap nut.
  • Page 75: Power Supply

    7 Connection FLUXUS F70x Power Supply Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). The external protective earth is connected to the equipotential bonding terminal on the housing of the transmitter FLUXUS F704**-A2, F705**-A2, F705**-NN (see Fig. 7.36 or Fig. 7.37). It always has to be connected. Attention! According to IEC 61010-1:2010, a switch has to be provided near the measuring instrument in the building installation, easily accessible for the user and marked as a disconnection device for the mea-...
  • Page 76 FLUXUS F70x 7 Connection ferrite nut cap nut compression part basic part power supply AGN ARS BGN BRS BR X_AV X_BV T1A T1B S1 T2A T3A T3B S3 T4A A+ B- P1+ P2+ P4+ P5a P6a P7a SA1 SA2 SA3 SA4 SB1 SB2 SB3 SB4 T1a T1b S2 T2a T3a T3b S4 T4a...
  • Page 77: Outputs

    7 Connection FLUXUS F70x Outputs Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). Attention! The outputs can only be connected to a low voltage circuit (max. 30 V AC or 42 V DC against earth). FLUXUS F704, F705, F706 For the connection of the output cable to the transmitter, see box Cable connection S.
  • Page 78 FLUXUS F70x 7 Connection Tab. 7.19: Circuits of the outputs output transmitter external circuit remark internal circuit connection active current < 500 Ω loop passive current loop = 4...24 V > 0.021 A [Ω] + 4 V example: = 12 V = 0...380 Ω...
  • Page 79 7 Connection FLUXUS F70x Tab. 7.19: Circuits of the outputs output transmitter external circuit remark internal circuit connection binary output (opto- ≤ 26 V relay) ≤ 100 mA F704 F705 F706 binary output (open = 5...24 V collector) Px+/Pxa [kΩ] = U [mA] = 1...4 mA P1...P4: R...
  • Page 80: Inputs

    FLUXUS F70x 7 Connection Inputs Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). Attention! The inputs can only be connected to a low voltage circuit (max. 30 V AC or 42 V DC against earth). inputs compression part basic part...
  • Page 81 7 Connection FLUXUS F70x 7.5.1 Temperature Input The temperature probes Pt100/Pt1000 (4-wire) can be connected to the inputs of the transmitter (optional) (see section 7.2). For the assignment and the activation of the temperature inputs see chapter 17. 7.5.2 Current Input An active or a passive current source can be connected to the current inputs of the transmitter (active current source: with own power supply, passive current source: with external power supply).
  • Page 82 FLUXUS F70x 7 Connection FLUXUS F709 • Prepare the input cable. • Connect the input cable to the terminals of the transmitter (see Fig. 7.4, Fig. 7.44 and Tab. 7.21). Tab. 7.21: Connection of a passive current source input transmitter external circuit comments internal circuit...
  • Page 83: Serial Interface

    7 Connection FLUXUS F70x Serial Interface Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS). 7.6.1 RS232 Interface The RS232 interface is located on the front plate of the transmitter (see Fig. 7.45, Fig. 7.46 or Fig. 7.47). •...
  • Page 84: Sensor Module (Sensprom)

    FLUXUS F70x 7 Connection FLUXUS F709 • Prepare the RS485 cable. • Connect the RS485 cable to the terminals of the transmitter (see Fig. 7.4, Fig. 7.48 and Tab. 7.19). C H A N N E L K L 1 K L 3 K L 5 K L 7...
  • Page 85 7 Connection FLUXUS F70x If transducers are replaced or added, the sensor module has to be replaced or added. Note! The serial number of sensor module and transducer have to be identical. A wrong or incorrectly con- nected sensor module will lead to incorrect measured values or to a measurement failure. FLUXUS F704, FLUXUS F705 •...
  • Page 86: Sd Card (Optional)

    FLUXUS F70x 7 Connection FLUXUS F709 • Disconnect the power supply from the transmitter. • For the connection of the AMP-Quick connector (connection system AS), see section 7.1.2.1. • For the connection of the separate sensor module (connection system TS): –...
  • Page 87: Start-Up Of The Transmitter

    FLUXUS F70x Start-up of the Transmitter Switching on As soon as the transmitter is connected to the power supply, the display indicates which FLEXIM FLUXUS transducer has been detected at which channel. F70x -XXXXXXXX Afterwards, the serial number of the transmitter is displayed for a short time.
  • Page 88 FLUXUS F70x 8 Start-up of the Transmitter 8.3.2 Program Branches • Program branch Parameter input of the pipe and medium parameters • Program branch Measuring processing of the steps for the measurement • Program branch Output Options setting of the physical quantity, the unit of measurement and the parameters for the measured value transmission •...
  • Page 89: Hotcodes

    8 Start-up of the Transmitter FLUXUS F70x 8.3.3 Navigation A vertical arrow ↕ will be displayed if the menu item contains a scroll list. The current list item will be displayed in the lower line. Use key to select a list item in the lower line. Press ENTER. Parameter ↕...
  • Page 90: Language Selection

    As soon as the measurement begins, all current measuring parameters will be stored in a non-volatile INIT-resistant EPROM. The measurement will be interrupted if the power supply fails. All input data will remain stored. After the return of the power supply, the serial number is displayed for a few seconds. FLEXIM FLUXUS F70x -XXXXXXXX The interrupted measurement is continued.
  • Page 91: Basic Measurement

    9 Basic Measurement FLUXUS F70x Basic Measurement Attention! Observe the Safety Instructions for the Use in Explosive Atmosphere (see document SIFLUXUS). The pipe and medium parameters are entered for the selected measuring point (see chapter 5). The parameter ranges are limited by the technical characteristics of the transducers and of the transmitter.
  • Page 92 FLUXUS F70x 9 Basic Measurement 9.1.3 Pipe Material The pipe material has to be selected to be able to determine the sound speed. The sound speed for the materials in the scroll list are stored in the transmitter. Select the pipe material. Pipe Material ↕...
  • Page 93: Input Of The Medium Parameters

    9 Basic Measurement FLUXUS F70x Input of the Medium Parameters Select the medium from the scroll list. Medium ↕ If the medium is not in the scroll list, select Other Medium. Press ENTER. Water It is possible to specify which media will be displayed in the scroll list (see section 13.5). For the programmed parameters of common media see annex C.3.
  • Page 94: Other Parameters

    Press ENTER. Note! If standard transducer parameters are used, FLEXIM cannot guarantee for the precision of the mea- sured values. A measurement might even be impossible. If Special Version has been selected, enter the 6 transducer parameters specified by Transd.
  • Page 95: Selection Of The Channels

    9 Basic Measurement FLUXUS F70x Selection of the Channels The channels on which a measurement is being made can be activated individually. Select program branch Measuring. Press ENTER. par>MEA<opt sf Measuring If this error message is displayed, the parameters are not complete. Enter the missing pa- par>MEA<opt sf rameters in the program branch Parameter.
  • Page 96 FLUXUS F70x 9 Basic Measurement 9.6.1 Fine Adjustment of the Transducer Distance 9 Basic Measurement If the displayed transducer distance is adjusted, press ENTER. Transd. Distance The measuring for the positioning of the transducers is started. mm ! The amplitude of the received signal is displayed by the bar graph S=. S=...
  • Page 97: Start Of The Measurement

    9 Basic Measurement FLUXUS F70x Tab. 9.1: Standard values for signal optimization transducer frequency difference between the optimum and the entered transducer distance [mm] (third character of the technical type) shear wave transducer lamb wave transducer -50...+100 -35...+60 -25...+40 -10...+20 -6...+10 -3...+5 Enter the new adjusted transducer distance.
  • Page 98: Displaying The Measured Values

    FLUXUS F70x 10 Displaying the Measured Values Displaying the Measured Values The physical quantity is set in the program branch Output Options (see section 10.1). During the measurement, the designation of the physical quantity is displayed in the upper line, the measured value in the lower line.
  • Page 99: Adjustment Of The Display

    10 Displaying the Measured Values FLUXUS F70x All channels The measured values of all channels (measuring and calculation channels) are displayed. The next active channel is dis- played after min. 1.5 s. Only calculation channels Only the measured values of the calculation channels are displayed. The next active calculation channel is displayed after min.
  • Page 100: Status Line

    FLUXUS F70x 10 Displaying the Measured Values 10.4 Status Line Important data on the ongoing measurement are displayed in the status line. The quality and precision of the ongoing measurement can be estimated. Press key during the measurement to scroll through the upper line to the status line. A: S3 Q9 c...
  • Page 101: Advanced Measuring Functions

    11 Advanced Measuring Functions FLUXUS F70x Advanced Measuring Functions 11.1 Damping Factor Each displayed measured value is a floating average of all measured values of the last x seconds, with x being the damp- ing factor. A damping factor of 1 s means that the measured values are not averaged because the measuring rate is ap- prox 1/s.
  • Page 102: Settings Of The Hybridtrek Mode

    FLUXUS F70x 11 Advanced Measuring Functions When the measurement is stopped The behavior of the totalizers when the measurement is stopped or after a RESET of the transmitter is set in Special Funct.\SYSTEM settings\Measuring\Quantity recall. If on is selected, the values of the totalizers will be stored and used for the next measure- Quantity recall ment.
  • Page 103: Upper Limit Of The Flow Velocity

    11 Advanced Measuring Functions FLUXUS F70x If the automatic toggling between the TransitTime and the NoiseTrek mode is activated, the TT-Failed |After toggling parameters have to be configured. NoiseTrek | 40s  Enter the time after which the transmitter has to toggle to the NoiseTrek mode if there are no valid measured values in the TransitTime mode.
  • Page 104: Cut-Off Flow

    FLUXUS F70x 11 Advanced Measuring Functions 11.5 Cut-off Flow The cut-off flow is a lower limit for the flow velocity. All measured flow velocities that are lower than the limit and their de- rived values are set to zero. The cut-off flow can depend on the flow direction or not. The cut-off flow is set in Special Funct.\SYSTEM set- tings\Measuring\Cut-off Flow.
  • Page 105: Measurement Of Highly Dynamic Flows (Fastfood Mode)

    11 Advanced Measuring Functions FLUXUS F70x 11.7 Measurement of Highly Dynamic Flows (FastFood Mode) The FastFood mode enables the measurement of flows with high dynamics. A continuous adaptation to changing measuring conditions which takes place in the normal measuring mode is only par- tially realized in the FastFood mode.
  • Page 106 FLUXUS F70x 11 Advanced Measuring Functions 11.8.1 Characteristics of the Calculation Channels In the program branch Parameter, the measuring channels to be used for the calculation and the calculation function have to be entered. A calculation channel cannot be attenuated. The damping factor has to be set separately for each measuring channels. Two cut-off flow values for each calculation channel can be defined.
  • Page 107 11 Advanced Measuring Functions FLUXUS F70x 11.8.3 Output Options for a Calculation Channel Select a calculation channel in the program branch Output Options. Press ENTER. Output Options ↕ for Channel Select the physical quantity to be calculated. Press ENTER. Physic. Quant. ↕ Mass Flow Make sure that the physical quantity selected for the calculation channel can be calculated from the physical quantities of the selected measuring channels.
  • Page 108: Change Of The Limit For The Inner Pipe Diameter

    FLUXUS F70x 11 Advanced Measuring Functions 11.8.4 Measurement with Calculation Channels Select program branch Measuring. Press ENTER. par >MEA< opt sf Measuring Activate the necessary channels. Calculation channels are activated or deactivated in the CHANN: A B >Y< Z same way as the measuring channels. Press ENTER. MEASUR ...
  • Page 109 11 Advanced Measuring Functions FLUXUS F70x 11.10.1 Configuration of the Remote Function The remote function can be configured for each individual channel. Select Special Funct.\SYSTEM settings\Remote inputs. Press ENTER. SYSTEM settings↕ Remote inputs Select the channel for which a remote function is to be activated. Press ENTER. Remote inputs ↕...
  • Page 110 FLUXUS F70x 11 Advanced Measuring Functions Activating the FastFood mode Select the list item FastFoodMode. Press ENTER. A: Remote-Func. ↕ The FastFood mode is activated for the duration of the switching signal. FastFoodMode The FastFood mode is deactivated when the voltage at the binary input is switched back to the passive range.
  • Page 111: Program Code

    11 Advanced Measuring Functions FLUXUS F70x 11.11 Program Code An ongoing measurement can be protected from an inadvertent intervention by means of a program code. If a program code has been defined, it will be requested when there is an intervention in the measurement (key BRK). 11.11.1 Defining a Program Code Select Special Funct.\Program code.
  • Page 112: Data Logger And Transmission Of Data

    FLUXUS F70x 12 Data Logger and Transmission of Data Data Logger and Transmission of Data The transmitter has a data logger in which the measured values are stored during the measurement (see section 12.1). In addition, the measurement data can be transmitted to a PC via the serial interface or stored on an SD card (see section 12.2).
  • Page 113 12 Data Logger and Transmission of Data FLUXUS F70x 12.1.3 Settings of the Data Logger Select program branch Special Funct.\SYSTEM settings\Storing. It contains the following menu items: • ringbuffer • storage mode • storing of the totalizers • storing of the signal amplitude •...
  • Page 114 FLUXUS F70x 12 Data Logger and Transmission of Data Start of the storing If it is necessary to synchronize the storing of measured values on several transmitters, the starting time of the storing can be set. Select the starting time of the storing of measured values. Start logger ↕...
  • Page 115: Transmission Of Data

    12 Data Logger and Transmission of Data FLUXUS F70x Max. 100 series of measured values can be stored. The number of series of measured values depends on the total num- ber of measured values stored in the previous series of measured values. The time at which the data logger memory will be full can be displayed during the measurement.
  • Page 116 FLUXUS F70x 12 Data Logger and Transmission of Data 12.2.3 Format of the Measurement Data Select Special Funct.\SYSTEM settings\serial transmis.. Select on if the space characters are not to be transmitted. Press ENTER. SER:kill spaces The file size will be considerably smaller (shorter transmission time). >ON<...
  • Page 117 12 Data Logger and Transmission of Data FLUXUS F70x 12.2.5 Online Transmission of Data to a Terminal Program or an SD Card If data are transmitted online to a terminal program: • Start the terminal program. • Enter the transmission parameters into the terminal program (see section 12.2.4). The transmission parameters of the terminal program and of the transmitter have to be identical.
  • Page 118 12 Data Logger and Transmission of Data 12.2.7 Offline Transmission of Data with the Program FluxData The measurement data in the data logger are transmitted to a PC via the serial interface RS232 with the FLEXIM program FluxData. Settings in the program Start the program FluxData V3.0 or higher on the PC.
  • Page 119 12 Data Logger and Transmission of Data FLUXUS F70x Stop of the transmission of data Select the menu: File > Save. Select the series of measurement to be stored. Click on OK. Select the path on which the data should be stored. Enter the file name.
  • Page 120 FLUXUS F70x 12 Data Logger and Transmission of Data The line \DATA will be transmitted next. Afterwards the column titles will be transmitted for the respective channel (see Tab. 12.3) The measured values are transmitted afterwards. Example: \DATA \*MEASURE; Q_POS; Q_NEG;...
  • Page 121: Libraries

    13 Libraries FLUXUS F70x Libraries The internal material database of the transmitter contains parameters for pipe and lining materials as well as for media. It can be extended with user-defined materials or media. User-defined materials and media will always be displayed in the scroll lists of the program branch Parameter.
  • Page 122: Input Of Material/Medium Parameters Without The Extended Library

    FLUXUS F70x 13 Libraries The occupancy of the coefficient memory is displayed for a few seconds. USER AREA: used Select yes to start the partitioning. Press ENTER. Format NOW? >YES< The coefficient memory is partitioned accordingly. This procedure takes a few seconds. FORMATTING ...
  • Page 123: Extended Library

    13 Libraries FLUXUS F70x Material parameters Enter the sound speed of the material. Press ENTER. c-Material For the sound speed of some materials annex C.1. 1590.0 Enter the roughness of the material. Press ENTER. Roughness For the typical roughness of some materials see annex C.2. Medium parameters Enter the average sound speed of the medium.
  • Page 124 If, however, the process conditions fluctuate strongly and the medium parameters depend strongly on the temperature (e.g., viscosity of a hydraulic oil), polynomials or customized interpolation functions should be used. Contact FLEXIM to find the best solution for the measuring task.
  • Page 125 13 Libraries FLUXUS F70x Select edit to edit the material/medium parameters or delete to delete the material/me- USER MATERIAL dium and to return to the scroll list Edit Material or Edit Medium. >EDIT< delete This display will only be indicated if an already existing material/medium has been select- Enter the designation of the material/medium.
  • Page 126: Deleting A User-Defined Material/Medium

    FLUXUS F70x 13 Libraries 13.3.4 Input of Heat Flow Coefficients Note! The heat flow coefficients can also be edited with the programs FluxData and FluxKoef. Note! The entered coefficients will not be checked. Absurd values can result in wrong measured values or in permanent system errors.
  • Page 127: Arrangement Of The Material/Medium Scroll List

    13 Libraries FLUXUS F70x 13.5 Arrangement of the Material/Medium Scroll List The materials and media to be displayed in the program branch Parameter are arranged in the material scroll list and in the medium scroll list. Note! User-defined materials/media will always be displayed in the scroll lists of the program branch Pa- rameter.
  • Page 128 FLUXUS F70x 13 Libraries 13.5.3 Adding all Materials/Media to the Scroll List Select Add all to add all materials/media of the database to the current scroll list. Press Material list ↕ ENTER. >Add all 13.5.4 Removing a Material/Medium from the Scroll List Select Remove Material or Remove Medium to remove a material/medium from the Material list ↕...
  • Page 129: Settings

    14 Settings FLUXUS F70x Settings 14.1 Time and Date The transmitter has a battery-powered clock. Measured values are automatically stored with the date and time. 14.1.1 Time Select Special Funct.\SYSTEM settings\Set Clock. Press ENTER. SYSTEM settings↕ Set Clock The current time is displayed. Select ok to confirm the time or new to set the time. Press TIME 11:00 ENTER.
  • Page 130 FLUXUS F70x 14 Settings During the next scroll through the program branch Parameter, the outer pipe diameter Outer Diameter that corresponds to the entered pipe circumference will be displayed. 57.3 example: 180 mm :  = 57.3 mm Note! The pipe circumference is only edited temporarily. When the transmitter switches back to the display of the pipe circumference (internal recalculation), slight rounding errors may occur.
  • Page 131 14 Settings FLUXUS F70x 14.2.7 Error Value Delay The error value delay is the time after which an error value will be sent to an output if no valid measured values are avail- able. Select edit to enter an error value delay. Select damping if the damping factor is to be Error-val.
  • Page 132: Measurement Settings

    FLUXUS F70x 14 Settings 14.3 Measurement Settings Select Special Funct.\SYSTEM settings\Measuring. Press ENTER. SYSTEM settings↕ Measuring Note! The settings of the menu item Measuring will be stored at the end of the dialog. If the menu item is quit before the end of the dialog, the settings will not be effective. This menu item will only be displayed if a WaveInjector is in the scope of supply (see user WaveInjector manual of the WaveInjector).
  • Page 133: Setting The Contrast

    14 Settings FLUXUS F70x 14.4 Setting the Contrast Select Special Funct.\SYSTEM settings\Miscellaneous to set the contrast of SYSTEM settings↕ the display of the transmitter. Press ENTER. Miscellaneous The contrast of the display is adjusted with the following keys: SETUP DISPLAY increases the contrast ←...
  • Page 134: Superuser Mode

    FLUXUS F70x 15 SuperUser Mode SuperUser Mode The SuperUser mode offers the possibility of an advanced analysis of the signal and the measured values as well as the definition of additional parameters adapted to the measuring point, in order to achieve better measuring values or during experimental work.
  • Page 135 15 SuperUser Mode FLUXUS F70x 15.3.1 Profile Bounds Select user if the profile bounds are to be defined. If factory is selected, the default pro- A:Profile bounds file bounds will be used and the menu item Calibration will be displayed (see section factory >USER<...
  • Page 136: Limit Of The Signal Amplification

    FLUXUS F70x 15 SuperUser Mode Example 1: Slope: 1.1 Offset: -10.0 cm/s = -0.1 m/s If a flow velocity v = 5 m/s is measured, before the calculation of the derived quantities, it will be cor- rected as follows: = 1.1 5 m/s - 0.1 m/s = 5.4 m/s Example 2: Slope: -1.0...
  • Page 137: Detection Of Long Measurement Failures

    15 SuperUser Mode FLUXUS F70x Example: fixed upper value of the sound speed thresh.: 2 007 m/s offset: 600 m/s value of the sound speed curve at the operating point: 1 546 m/s As 1 546 m/s + 600 m/s = 2 146 m/s is greater than the fixed upper value 2 007, this value will be used as the upper limit of the sound speed when the plausibility of the signal is evaluated.
  • Page 138: Temperature-Based Heat Flow Cut-Off

    FLUXUS F70x 15 SuperUser Mode Total digits = Fixed to x digit The number of decimal points is constant. The max value of the totalizer is reduced with each additional decimal place. decimal places max. value max. display < 10 ±9999999999 <...
  • Page 139: Display Of The Last Valid Measured Value

    15 SuperUser Mode FLUXUS F70x 15.11 Display of the Last Valid Measured Value If the signal is not sufficient for a measurement, usually UNDEF will be displayed. Instead of UNDEF, it is also possible to display the last valid measured value. Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous.
  • Page 140: Heat Flow Measurement

    FLUXUS F70x 16 Heat Flow Measurement Heat Flow Measurement If the transmitter has the optional heat quantity measurement and two temperature inputs, the heat flow can be measured. A temperature probe is fixed on the supply and the return line. For the mounting of the temperature probes see section 6.5.
  • Page 141: Calculation Of The Heat Flow Rate

    16 Heat Flow Measurement FLUXUS F70x 16.1 Calculation of the Heat Flow Rate The heat flow rate is calculated by the following formula: · Φ = k with Φ – heat flow rate – heat coefficient · – volumetric flow rate –...
  • Page 142: Btu Mode

    FLUXUS F70x 16 Heat Flow Measurement 16.2.3 Input of a Constant Temperature If the supply or return temperature is known and constant during the whole measurement, this temperature can be entered in the transmitter. Note! A constant temperature should be entered if, e.g., the supply temperature can only be measured with difficulty but is known and constant.
  • Page 143 16 Heat Flow Measurement FLUXUS F70x 16.3.2 Assignment of the Transducers and the Temperature Inputs The position of the transducers and the temperature inputs can be assigned in accordance with the application. Select Special Funct.\SYSTEM settings\Proc. inputs\Link temperature. In case of a heating application, select heat, in case of a cooling application, select A:Thermal energy chill.
  • Page 144: Measurement

    FLUXUS F70x 16 Heat Flow Measurement In the BTU mode, additional units of measurement are available for the physical quantity and the heat quantity (see sec- tion 11.2). The unit of measurement displayed during the measurement will be adjusted automatically: unit of measurement unit of measurement of the heat flow...
  • Page 145: Two Independent Heat Flow Rate Measurements

    16 Heat Flow Measurement FLUXUS F70x 16.5 Two Independent Heat Flow Rate Measurements If the transmitter has 2 measuring channels and 2 temperature inputs for each measuring channel, it is possible to conduct 2 independent heat flow measurements at the same time. Tab. 16.1 shows a typical configuration of the temperature in- puts.
  • Page 146 FLUXUS F70x 16 Heat Flow Measurement 16.6.2 Display of the State of Aggregation During the heat flow measurement, the state of aggregation of the medium can be displayed in the upper line by pressing display explanation state of aggregation in the supply line state of aggregation in the return line The medium is completely gaseous.
  • Page 147: Inputs

    17 Inputs FLUXUS F70x Inputs 17.1 Binary Inputs The transmitter can be equipped with 1 or 2 binary inputs. Using the binary outputs, it is possible to remotely trigger some functions of the transmitter (see section 11.10). 17.2 Analog Inputs External sensors can be connected to the inputs (optional) to measure the following physical quantities: •...
  • Page 148 FLUXUS F70x 17 Inputs 17.2.2 Selection of the Temperature Probe Select Special Funct.\SYSTEM settings\Proc. inputs. Press ENTER. SYSTEM settings↕ Proc. inputs Select the list item PT100/PT1000. Proc. inputs ↕ PT100 / PT1000 Select the temperature probe. Input T1 ↕ >PT100< pt1000 If necessary, select the temperature probe for Input T2...T4 accordingly.
  • Page 149 17 Inputs FLUXUS F70x 17.2.4.2 Activation of Other Inputs Attention! Observe the correct polarity to avoid damaging the current source. A permanent short circuit can lead to the destruction of the current input. Inputs have to be activated if the measured values are to be displayed, stored and/or transmitted together with the other measured values.
  • Page 150: Outputs

    FLUXUS F70x 18 Outputs Outputs If the transmitter is equipped with outputs, they have to be installed and activated before they can be used: • assign a measuring channel (source channel) to the output (if the transmitter has more than one measuring channel) •...
  • Page 151 18 Outputs FLUXUS F70x Tab. 18.1: Configuration of the outputs source item list item output physical quantity selected in the program branch Output Options Measuring value actual measure flow, independently of the physical quantity selected in the program Flow branch Output Options heat flow, independently of the physical quantity selected in the program Heatflow branch Output Options...
  • Page 152 FLUXUS F70x 18 Outputs 18.1.1 Output Range During the configuration of an analog output, the output range is defined. Select a list item I1 Output range↕ or other range... to enter the output range manually. 4/20 mA If other range... has been selected, enter the values Output MIN and Output MAX. I1 Output MIN ↕...
  • Page 153 18 Outputs FLUXUS F70x Tab. 18.3: Examples for the error output list item for the error output output signal I [ m A ] Error-value ↕ Minimum (4.0mA) Error-value ↕ I [ m A ] Hold last value Error-value ↕ I [ m A ] Maximum (20.0mA) Error-value...
  • Page 154: Error Value Delay

    FLUXUS F70x 18 Outputs 18.1.3 Function Test The function of the installed output can now be tested. Connect a multimeter with the installed output. Test of the Analog Outputs The current output is tested in the display. Enter a test value. It has to be within the output I1:Output Test range.
  • Page 155: Configuration Of A Frequency Output As A Pulse Output

    18 Outputs FLUXUS F70x 18.3.1 Measuring Range of the Analog Outputs After an analog output has been activated in the program branch Output Options, the measuring range of the source item has to be entered. Select sign if the sign of the measured values is to be considered for the output. Meas.Values Select absolut if the sign is not to be considered.
  • Page 156: Activation Of A Binary Output As A Pulse Output

    FLUXUS F70x 18 Outputs Select in the scroll list the measuring channel to be assigned to the output as the source F1 Source chan.↕ channel. Press ENTER. Channel A: Select in the scroll list Measuring value (but not Impuls!). Press ENTER. F1 Source item ↕...
  • Page 157: Activation Of A Binary Output As An Alarm Output

    18 Outputs FLUXUS F70x 18.6 Activation of a Binary Output as an Alarm Output Note! The menu item Alarm Output will only be displayed in the program branch Output Options if an alarm output has been installed. Select in the program branch Output Options the channel for which an alarm output is Output Options ↕...
  • Page 158 FLUXUS F70x 18 Outputs 18.6.2 Setting the Limits If the switching condition MAX or MIN is selected in the scroll list func, the limit of the output will have to be defined: Select in the scroll list Input the physical quantity to be used for the comparison. The fol- R1 Input: ↕...
  • Page 159: Behavior Of The Alarm Outputs

    18 Outputs FLUXUS F70x 18.6.3 Defining the Hysteresis A hysteresis can be defined for the alarm output R1 to prevent a constant triggering of the alarm due to small fluctuations of the measured values around the limit. The hysteresis is a symmetrical range around the limit. The alarm will be activated if the measured values exceed the up- per limit and deactivated if the measured values fall below the lower limit.
  • Page 160 FLUXUS F70x 18 Outputs 18.7.4 Alarm Outputs During the Measurement An alarm output with switching condition MAX or MIN will be updated max. once per second to avoid humming (i.e. fluctua- tion of the measured values around the value of the switching condition). An alarm output of the type NON-HOLD will be activated if the switching condition is met.
  • Page 161: Deactivation Of The Outputs

    18 Outputs FLUXUS F70x Tab. 18.6: Pictograms for the alarm state indication current state func mode (switching condition) (holding behavior) (switching function) closed NON-HOLD NO Cont. open HOLD NC Cont. +→- -→+ QUANT. ERROR Example: R1 = 18.8 Deactivation of the Outputs If the programmed outputs are no longer required, they can be deactivated.
  • Page 162: Troubleshooting

    A program code has been defined. Press key C and enter the program code. The backlight of the display does not work, but all other functions are available. The backlight is defective. This problem does not affect the other functions of the display. Send the transmitter to FLEXIM for repair.
  • Page 163: Problems With The Measurement

    • If no exclamation mark "!" is displayed, a measurement at the selected measuring point is not possible. Loss of signal during the measurement • If the pipe had run empty: Was there no measuring signal afterwards? Contact FLEXIM. • Wait briefly until acoustic contact is reestablished. The measurement can be interrupted by a temporarily higher propor- tion of gas bubbles and solids in the medium.
  • Page 164: Maximum Acoustic Contact

    The flow is in the transition range between laminar and turbulent flow where flow measurement is difficult Calculate the Reynolds number of the flow at the measuring point with the program FluxFlow (free download: www.flex- im.com). Contact FLEXIM. 19.5 Large Deviations of the Measured Values...
  • Page 165: Problems With The Totalizers

    19 Troubleshooting FLUXUS F70x 19.6 Problems with the Totalizers The values of the totalizer are too high See Special Function\SYSTEM settings\Measuring\Quantity recall. If this menu item is activated, the val- ues of the totalizer will be stored. The totalizer will continue with this value at the start of the next measurement. The values of the totalizer are too low One of the totalizers has reached the upper limit and has to be reset to zero manually.
  • Page 166: A Menu Structure

    FLUXUS F70x A Menu Structure Menu Structure INIT- resistant Program Branch Parameter main menu: selection of the program branch Parameter >PAR< mea opt sf Parameter selection of a measuring channel (A, B) or of a calculation channel (Y, Z) Parameter ↕...
  • Page 167 A Menu Structure FLUXUS F70x INIT- resistant selection of the range of the sound speed c-Medium range auto: The area around the average sound speed is defined by the transmitter. auto >USER< user: The area around the average sound speed has to be entered. input of the range around the average sound speed of the medium c-Medium=1500m/s This display will only be indicated if Other Medium has been selected.
  • Page 168 FLUXUS F70x A Menu Structure INIT- resistant input of the number of sound paths A: Sound Path display of the transducer distance to be adjusted between the inner edges of the Transd. Distance transducers A:54 mm Reflex Program Branch Output Options main menu: selection of the program branch Output Options par mea >OPT<...
  • Page 169 A Menu Structure FLUXUS F70x INIT- resistant input of the lowest/highest measured value to be expected for the current output Zero-Scale Val. The values are assigned to the lower/upper limit of the output range. 0.00 m3/h These displays will only be indicated if Current Loop is activated. Full-Scale Val.
  • Page 170 FLUXUS F70x A Menu Structure INIT- resistant SYSTEM settings selection of Special Funct.\SYSTEM settings Special Funct. ↕ SYSTEM settings SYSTEM settings\Set Clock selection of the displays for the input of the date and the time SYSTEM settings↕ Set Clock SYSTEM settings\Libraries selection of the displays for the management of the material and medium scroll SYSTEM settings↕...
  • Page 171 A Menu Structure FLUXUS F70x INIT- resistant SYSTEM settings\Libraries\Extended Library selection of the displays for the activation of the extended library Libraries ↕ Extended Library activation of the extended library Extended Library >ON< SYSTEM settings\Dialogs/Menus selection of the displays for the activation/deactivation or setting of the menu SYSTEM settings↕...
  • Page 172 FLUXUS F70x A Menu Structure INIT- resistant selection of the unit of measurement for the pressure Pressure >[bar]< [psi] selection if lb/ft is to be used as the unit of measurement for the density Density [lb/ft3] >YES< selection of the unit of measurement for the density Density unit This display will only be indicated if lb/ft has not been selected as the unit of...
  • Page 173 A Menu Structure FLUXUS F70x INIT- resistant enabling of the NoiseTrek mode Enable NoiseTrek >ON< Selection if the toggling between the TransitTime and the NoiseTrek mode has Auto NoiseTrek ? to be carried out manually or automatically. >YES< This display will only be indicated if the NoiseTrek mode is enabled. Input of the time after which the transmitter has to toggle to the NoiseTrek mode TT-Failed |After...
  • Page 174 FLUXUS F70x A Menu Structure INIT- resistant Input of the cut-off flow for negative measured values -Cut-off Flow range: -12.7...0 cm/s -2.5 cm/s default: -2.5 cm/s This display will only be indicated if Cut-off Flow = sign and Cut-off Flow = user has been selected. Input of the cut-off flow for the absolute value of the measured values Cut-off Flow range: 0...12.7 cm/s...
  • Page 175 A Menu Structure FLUXUS F70x INIT- resistant Input of the limit for the temperature difference. Thermal flow ->0 All temperature differences between the supply and return line that are lower if |dT|< 0.0 C than this value will be set to zero. range: 0...5.0 °C 0: no temperature-based heat flow cut-off This display will only be indicated if the SuperUser mode is activated and Ther-...
  • Page 176 FLUXUS F70x A Menu Structure INIT- resistant input of the offset. A:Offset= range: -12.7...+12.7 cm/s cm/s 0: no offset This display will only be indicated if the SuperUser mode is activated and Cali- bration = on is selected. SYSTEM settings\Proc. outputs selection of the displays for the setting of the outputs of the transmitter SYSTEM settings↕...
  • Page 177 A Menu Structure FLUXUS F70x INIT- resistant selection of the character for column separation SER:col-separat. ’;’ >’TAB’< selection of the serial interface Send Offline via default: RS232 RS232 >RS485< This display will only be indicated if the transmitter has an RS485 interface. SYSTEM settings\Network change of the settings of the transmission parameters of the RS485 interface SYSTEM settings↕...
  • Page 178 FLUXUS F70x A Menu Structure INIT- resistant Delete Meas.Val. selection of the displays for the deleting of stored measured values Special Funct. ↕ Delete Meas.Val. confirmation for the deleting of measured values Really Delete? This display will only be indicated if measured values are stored in the data log- >YES<...
  • Page 179 A Menu Structure FLUXUS F70x INIT- resistant input of the roughness of the material Roughness confirmation that the changes are to be stored Save changes This display will only be indicated if a new material has been entered or the >YES<...
  • Page 180 FLUXUS F70x A Menu Structure INIT- resistant input of the kinematic viscosity of the medium VISCOSITY mm2/s input of the operating density of the medium DENSITY g/cm3 confirmation that the changes are to be stored Save changes This display will only be indicated if a new medium has been entered or the >YES<...
  • Page 181: B Units Of Measurement

    B Units of Measurement FLUXUS F70x Units of Measurement Length/roughness Temperature unit of measurement description unit of measurement description millimeter degree Celsius °C inch degree Fahrenheit inch °F Pressure unit of measurement description bar (absolute) bar(a) bar (relative) bar(g) pound per square inch (absolute) psi(a) pound per square inch (relative) psi(g)
  • Page 182 FLUXUS F70x B Units of Measurement Volumetric flow rate Volume (totalized) unit of measurement description unit of measurement cubic meter per day m3/d cubic meter per hour m3/h cubic meter per minute m3/min cubic meter per second m3/s 1000 cubic meters per hour km3/h milliliter per minute l or m3*...
  • Page 183 B Units of Measurement FLUXUS F70x Mass flow rate Mass (totalized) unit of measurement description unit of measurement metric ton per hour metric ton per day kilogram per hour kg/h kilogram per minute kg/min kilogram per second kg/s gram per second pound per day lb/d pound per hour...
  • Page 184 FLUXUS F70x B Units of Measurement Flow Nomogram (Metrical) volumetric flow rate volumetric flow rate V o l u m e n f l u s s V o l u m e n f l u s s 1 0 0 1 0 0 1 0 0 1 0 0...
  • Page 185 B Units of Measurement FLUXUS F70x Flow Nomogram (Imperial) volumetric flow rate volumetric flow rate V o l u m e n f l u s s V o l u m e n f l u s s 1 0 0 1 0 0 1 0 0 1 0 0...
  • Page 186: C Reference

    The following tables provide assistance for the user. The accuracy of the data depends on the composition, temperature and processing of the material. FLEXIM does not assume liability for any inaccuracies. Sound Speed of Selected Pipe and Lining Materials at 20 °C The values of some of these materials are stored in the internal database of the transmitter.
  • Page 187 C Reference FLUXUS F70x Typical Roughnesses of Pipes The values are based on experience and measurements. material absolute roughness [mm] drawn pipes of non-ferrous metal, glass, plastics and light metal 0…0.0015 drawn steel pipes 0.01…0.05 fine-planed, polished surface max. 0.01 planed surface 0.01…0.04 rough-planed surface...
  • Page 188 FLUXUS F70x C Reference medium (display) explanation sound speed kinematic viscosity density [m/s] [mm²/s] [g/cm³] ISO VG 32 1 487 78.0 0.8690 ISO VG 32 ISO VG 46 1 487 126.7 0.8730 ISO VG 46 ISO VG 68 1 487 201.8 0.8750 ISO VG 68...
  • Page 189 C Reference FLUXUS F70x Properties of Water at 1 bar and at Saturation Pressure medium temperature medium pressure density specific heat* [°C] [bar] [kg/m³ ] [kJ/kg/K ¹] 999.8 4.218 999.7 4.192 998.3 4.182 995.7 4.178 992.3 4.178 988.0 4.181 983.2 4.184 977.7 4.190...
  • Page 190: D Declarations Of Conformity

    FLUXUS F70x D Declarations of Conformity Declarations of Conformity UMFLUXUS_F7V4-6-2EN, 2017-10-01...

Table of Contents