BIEGLER Instructions for use - BW 585 14 TECHNICAL DATA Device: Blood and infusion warmer IMPORTANT: Type designation: BW 585 These instructions are an essential part of the device. They must therefore be Voltage: 230 V / 50-60 Hz or 110 V / 50 Hz...
Instructions for use - BW 585 BIEGLER BIEGLER Instructions for use - BW 585 12 SYMBOLS Table of contents 1 Warnings and safety instructions..............2 Certifies compliance with 93/42/EEC 0123 2 Description.......................4 2.1 General Description..................4 Observe the instructions for use 2.2 Scope of delivery..................4 Do not dispose of this product as unsorted municipal waste 2.3 Consumable material...................4...
RF communications equipment and the BW 585 been no impairment. The BW 585 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the BW 585 can help prevent electromagnetic interference by If the high temperatur alarm is triggered the supply of liquid to the patient must •...
Page 5
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the BW 585 is used exceeds the applicable RF compliance level above, the BW 585 should be observed to verify normal operation.
Table 201 – Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS (see 6.8.3.201 a) 3)) The BW 585 is intended for use in the electromagnetic environment specified below. The customer or the user 3.1 SETTING UP PROCEDURE of the BW 585 should assure that it is used in such an environment.
Prepare infusion or transfusion. Attention: The length of the tube between the device from and to the place of use. BW 585 and the patient must be at least 40 cm and the tube must not be streched. The manufacturer and/or supplier under no circumstances assume liability for slight negligence.
• with the applicable national standards, or if power supply during the period of The BW 585 is designed as a low-maintenance device. To preserve the quality use of the device is not guaranteed and functional safety, please comply with the following points: original spare parts material are not used or the maintenance interval is not •...
Page 10
CHECKING THE WARM-UP PERIOD The device should be checked for mechanical damage (general condition) and This is the time taken by the BW 585 to heat up to 41.0°C from room for the completeness of sticker information, particularly the plate on the reverse.