• #H˙Hexadecimal
• #O˙Octal
• #B˙Binary
Following are examples of invalid syntax command strings that will produce
errors:

Table 4.1 Syntax Error Examples

COMMAND
Mode dB
Mode:dBm Range:Auto
DIS ?
Ran3;dis?
Disply ON
Wave
See Error Messages on page 71 for a list of error codes.

GPIB Command Reference Summary

Below is a summary reference list of all GPIB commands that the FPM-8200
responds to:

Table 4.2 GPIB Summary Reference Command List

Syntax
*CAL?
CAL:USER <nrf value>
CAL:USER?
*CLS
COND?
DELAY <nrf value>
DISplay <nrf value>
DISplay?
ENABle:COND <nrf
value>
ENABle:COND?
ENABle:EVEnt <nrf
value>
ENABle:EVEnt?
01_10
G P I B C O M M A N D R E F E R E N C E
GPIB Command Reference Summary
COMMENT
Compound command, missing colon between MODE and DB.
Missing semicolon between commands.
Space not allowed before question mark in queries.
Missing space between Ran and parameter 3.
Optional letters must follow the correct order.
Wavelength parameter missing.
Meaning
Calibrate
User calibration
User calibration?
Clear status
Condition?
Delay
Display on/off
Display value?
Set condition status enable
register
Condition status
enable register?
Event status
enable register
Event status
enable register?
Function
Calibrate internal analog to digital (A/D)
converter
Set a gain factor
Request user−cal gain factor
Clear status event registers
Request the status condition register
Delays processing of further commands for a
defined time
Turn front panel display on or off
Request the value on the measurement display
Set the condition status enable register
Request the condition status enable register
Set the event status enable register
Request the event status enable register