5.6.4 Tag Number Configuration
Tag Number is edited by turning the external adjustment screw. Up to 8 alphanumeric characters for HART or
16 alphanumeric characters for BRAIN can be set.
Blinking
Change the first Character
5.6.5 Pressure Unit Configuration
Pressure unit for the below table can be changed as below. By turning the external adjustment screw, user can
scroll between the various available pressure units.
5.6.6 Pressure LRV/URV Configuration
Pressure LRV and URV can be set. The number for each digit is changed by turning the external adjustment
screw and set by pressing the push button. Please refer to 5.6.7 Damping Time Constant Configuration for
how to change the numerical value.
When the setting is out of the limit, an alarm will be generated.
Blinking
Go to the next Character
screw
Blinking
Blinking
screw
Set all other
characters in the
same way
push
Save ?/
Cancel ?
[Available pressure units]
Select by the push button
kPa
Torr
atm
MPa
hPa
mmHg
push
psi
In addition to the above units, mmAq@4degC,
mmAq@20degC, mmWG@4degC and mmWG@20degC
are available for BRAIN communication type.
<5. Operation>
Save ?/
Cancel ?
bar
inH
O@4degC(39.2degF)
2
mbar
inH
O@20degC(68degF)
2
g/cm
2
inHg
kg/cm
2
ftH
O@4degC(39.2degF)
2
Pa
ftH
O@20degC(68degF)
2
mmH
O@4degC(39.2degF)
2
mmH
O@20degC(68degF)
2
IM 01C25W01-01EN
5-11
F0517.ai
F0518.ai

Advertisement

Table of Contents
loading

Table of Contents