4.8

Rotating Transmitter Section

The transmitter section can be rotated to any
desired position. Note that there is a stopper which
prevents the transmitter from being rotated more
than 360°.
1) Using the Allen wrench, remove the two
setscrews securing the transmitter section to
the capsule assembly.
2) Rotate the transmitter section slowly to the
desired position.
3) Tighten the two setscrews to a torque of
1.5 N·m {15 kgf·cm}.
WARNING
In the case of the explosion-proof/flameproof
type transmitter, do not rotate the transmitter part
in the hazardous area while the transmitter is
energized.
IMPORTANT
Do not rotate the transmitter section more than
the above limit.
Stopper
Figure 4.16
Rotating Transmitter Section
Transmitter section
Set screw
F0416.ai
<4. Installation>
4.9
Changing Integral Indicator
Direction
IMPORTANT
• Always turn OFF power and shut off and
release pressures before disassembly.
• For changing the integral indicator direction,
the transmitter must be removed to a non-
hazardous area.
An integral indicator can be installed in the following
three directions, Refer to subsection 6.4 for
attaching and removing the integral indicator.
Figure 4.17
Integral Indicator Direction
4-10
F0417.ai
IM 01C25W01-01EN