3. Specifications
Output signal 4 ... 20 mA, model TR30-W
Basic configuration
Analogue output
Measuring error per
IEC 60770, 23 °C ±5 K
Linearisation
Linearisation error
Switch-on delay, electrical
Current signal for fault
signal
Sensor short-circuit
Load R
A
Effect of load
Power supply U
B
Max. permissible residual
ripple
Power supply input
Power supply effect
Electromagnetic
compatibility (EMC)
Temperature units
Info data
Configuration and
calibration data
Electrical connection
Readings in % refer to the measuring span
For a correct determination of the overall measuring error, both sensor and transmitter
measuring deviations have to be considered.
1) The temperature transmitter should therefore be protected from temperatures over 85 °C (185 °F).
2) For measuring spans smaller than 50 K additional 0.1 K
3) ±0.2 % for measuring ranges with a lower limit less than 0 °C
4) Use resistance thermometers with shielded cable, and ground the shield on at least one end of the lead, if the
cables are longer than 30 m or leave the building.
WIKA operating instructions model TR30
Measuring range 0 ... 150 °C, other measuring
ranges are adjustable
4 ... 20 mA, 2-wire
1 % (Transmitter)
2)
Linear with temperature per DIN EN 60751
±0.1 %
3)
< 10 ms
Configurable in accordance with NAMUR NE43
downscale ≤ 3.6 mA
Not configurable, generally NAMUR downscale
≤ 3.6 mA
R
≤ (U
- 9 V) / 0.023 A with R
A
B
±0.05 % / 100 Ω
DC 10 ... 35 V
10 % at 24 V / maximum 300 Ω Load
Protected against reverse polarity
±0.025 % / V
EN 61326 emission (Group 1, Class B) and
immunity (industrial application)
NAMUR NE21
Configurable °C, °F, K
TAG No., descriptor and message can be stored
in transmitter
Permanently stored in EEPROM
M12 x 1 circular connector (4-pin)
DIN angular connector form A (for cable with
Ø 6 ... 8 mm, cross section max. 1.5 mm²)
upscale ≥ 21.0 mA
in Ω and U
A
B
, and also per
4)
EN
in V
11