Invacare HomeFill Oxygen System User Manual page 30

Table of Contents

Advertisement

Invacare® HomeFill® Oxygen System
Compressor cycles on and Off
or
O
Below Normal
2
(yellow) and filling (green) lights are alternating
or
O
below normal (yellow) light goes On and Off
2
O
below normal (YELLOW) light stays On
2
30
Problem
1. Check the concentrator liter flow setting is at 3 L/min or
less.
If necessary, adjust the liter flow setting (refer to the
concentrator User manual). Wait at least 20 minutes for the
GREEN light to appear. If it does not appear, proceed to
step 2.
2. The concentrator may need service, contact your home
health care provider or Invacare.
1. Ensure the concentrator has warmed up for at least 20
minutes. If it has not been 20 minutes, turn the compressor
Off until the concentrator warms up. If the YELLOW light still
remains on proceed to step 2.
2. Inspect the connection between the compressor and the
concentrator for damage or pinching.
• If damaged, replace.
• If pinched, straighten tubing. Turn the compressor Off for
at least 10 seconds. Turn the compressor On. If the YELLOW
light still remains on after 10 minutes, proceed to step 3.
3. The concentrator may need service, contact your home
health care provider or Invacare.
Solution
1100873-M