Invacare® HomeFill® Oxygen System
1. Make sure the concentrator is On. Refer to the
concentrator User Manual.
The maximum oxygen concentrator flow rate
for patient use is listed in the table below.
Otherwise, the oxygen level to the compressor
will be <90% and the compressor will not fill the
cylinder.
Invacare Oxygen
Concentrator
Platinum 5 and XL
Platinum 10
Perfecto
2
Patients with a higher prescribed flow should
not refill a cylinder while receiving supplemental
oxygen treatment from the oxygen concentrator.
2. Set the concentrator flow rate to the patient to 3 L/min
or less for Platinum 5, XL, and Perfecto
less for Platinum 10. Refer to the concentrator User
Manual.
3. Make sure the cylinder is connected to the compressor.
Refer to 4.3 Connecting the Cylinder to the Compressor,
page 19.
4. Push the power switch A on the control panel to the
On ( I ) position.
22
Maximum Output Flow
for Patient Use
3 L/min
5 L/min
3 L/min
and 5 L/min or
2
5. Examine the control panel B. The following sequence
of events should occur:
a. 0-3 minutes - The O
C will be lit while the compressor warms up.
b. After three minutes - The FILLING (GREEN) light D
will be on while the cylinder is filling.
c. The FULL (GREEN) light E will be on when the
cylinder is finished filling. Proceed to STEP 6.
The O
BELOW NORMAL (YELLOW) light
2
may come on if the oxygen level from the
concentrator has not yet reached or drops below
90%. The compressor will begin or resume filling
when the oxygen level reaches 90%. If light
does not go Off within 10 minutes, refer to 7.2
Troubleshooting, page 29.
If the ATTENTION (RED) light and audible alarm
F are on refer to 7.2 Troubleshooting, page 29.
6. Remove the full cylinder. Refer to 4.4 Disconnecting the
Cylinder from the Compressor, page 20.
7. Push the power switch to the Off (O) position.
Below Normal (YELLOW) light
2
1100873-M