SATRON VC Optical Consistency Transmitter
ERROR AL the digital output will change its state
ERROR AL the digital output will change its state
when there is an error
WARNG AL the digital output will change its state
when there is a warning
ERWNG AL the digital output will change its state
when there is a error and/or warning
HOLD
when the digital input is ON the whole
unit will be in a hold until the input is
OFF
.
NONE
no function
OFT ACKN the digital input mode [ON] setting off a
timer to overfeed.
RECIPE+1 digital input status [ON]
RECIPE+2 increase the number of active pre-
scription for one (RECIPE 1) or two
(RECIPE 2) if the recipe is I / O SEL.
TRB ZERO the digital input mode [ON] to reset the
value of the consistency of the factory.
DATA LOG the digital input mode is [ON] storing
data logging is permitted if the DATA
LOG parameter is set to DIGITAL I /
O.
FLSH ON
the digital input mode [ON] to set the
fl ush mode to [ON].
FLSH OFF the digital input mode [ON] to set the
fl ush mode [OFF].
FLSH OVT digital output is for fl ushing guidance.
SOURCE:
Select the source to which the digital output will change its
state.
PV is the process value selected by the user.
(value which is behind "U" on the display).
MA is the 1st current loop
ST is the sensor temperature located 5 millimeter
behind the optical lens
RANGE-% this will show a 0 to 100 % value correlating
to 4...20mA.
ON DELAY:
On delays can be used to delay digital output state from OFF >
ON transitions. The time can be selected in seconds in the
range of 0...300s. By default the off delay is not used.
OF DELAY:
Off delays can be used to delay digital output state from ON
> OFF transitions. The time can be selected in seconds in the
range of 0...300s. By default the off delay is not used.
OF TIMER:
Overfeed timer limits the time that the digital output can be
continuously in ON state.
The time can be selected in seconds in the range of 1...60000s.
By default the overfeed timer is not used.
Note: overfeed timer does not function if digital output is over-
ridden by HOLD function, when performing a I/O test in the
DIAGNOST menu or with HART CPU Control/ DOOverride.
.
.
.
IO2 SOURCE:
IO2 SOURCE:
The source for 2 nd mA out (PV,ST, ET, ...).
IO2 LRV:
The lower range value for 2 nd mA out (4 mA).
.
IO2 URV:
The upper range value for 2 nd mA out (20 mA).
IO2 DAMPING:
The time constant for 2 nd mA out (0 ... 60 s).
3.2 NEW SAMPLE
The new sample menu
START:
Store a new sample to memory.
SAMPLE H
O:
2
Restore the water point to memory.
3.3 CALIBRATION
The calibration menu.
RECIPE:
The settings for active recipe.
OFFSET:
The offset correction for calibration (default 0.0)
GAIN:
The gain correction for calibration (default 1.0)
USER.PNTS:
The number of points for multipoint calibration.
POINT.CNT
points 1 ... 16.
Their point of entry is given a number in either the
keyboard (EDIT) or by saving the real-time mea-
surement (SAMPLE).
Point out the value of the pair (user selectable unit)
is given a number in the keypad.
See the section of this manual for an example of
tuning to get more information on the complete
re-calibration.
USER MODE:
Select the method of interpolation between the
points.
INTERPL
SPLINE
TEXT:
Select the user name for the recipe.
12
12
calibrated count the number of
Select a linear interpolation.
Select the spline curve with interpola
tion.
BCs220AV
Revision 2
15.10.2014