6) Write protected
The loop device is write protected. The LC110H can not alter the
write protect status of a device. The command was ignored by the
loop device. HART response code 7.
7) Set to nearest possible value
The data value sent with the HART command was rounded or
truncated due to limitations within the loop device. The command
was accepted by the loop device. The resultant data values should
be checked. HART response code 8.
8) Update in progress
The loop device is updating the data value on a delayed basis. The
command was accepted by the loop device. HART response code 8.
9) Incorrect loop mA mode or value
The loop device is not in fixed current mode, or the mA value
supplied is invalid. The command was ignored by the loop device.
HART response code 9.
10) LRV too high
The LRV value received is greater than the Upper Transducer Limit of
this loop device, or some other physical device limitation is
exceeded. The command was ignored by the loop device. HART
response code 9.
11) Applied process too high
The data value received is greater than some device limit. The
command was ignored by the loop device. HART response code 9.
12) LRV too low
The LRV value received is less than the Lower Transducer Limit of
this loop device, or some other device physical limitation is
exceeded. The command was ignored by the loop device. HART
response code 10.
13) Applied process too low
The data value received is less than some device limit. The
command was ignored by the loop device. HART response code 10.