Martel LC-110 Reference Manual page 20

Table of Contents

Advertisement

Find Quality Products Online at:
6.1.2 In Circuit, LC-110H Loop Power
In "mA Measure with 24V" mode the LC-110H is in circuit and the
loop power is supplied by the LC-110H as shown below.
)
( +
)
( –
LC-110H
Precision Loop
Calibrator w/HART™
Measure 24V
0.00%
4.000mA
250Ω
MARTEL ELECTRONICS
100%
25%
0%
MENU
EXIT
Note: In this situation the 250Ω HART resistor is enabled. If your loop
already has 250Ω, do not enable the HART resistor.
6.1.3 Across Circuit, Communicator Only
In "Communicator Only" mode the LC-110H is across the circuit and
the loop power is supplied externally as shown below.
)
( +
24V
4 to 20mA
SUPPLY
)
( –
Note: In the Communicator Only Mode, there must be 250Ωs
resistance present in the loop.
18
GlobalTestSupply
www.
RL≥250Ω
LC-110H
Precision Loop
Calibrator w/HART™
PT12234
Device Setup & Data
Write LRV & URV Value
Trim, Set, & Zero
Device Diagnostic
Data Log & Config Log
MARTEL ELECTRONICS
100%
25%
0%
MENU
EXIT
.com
sales@GlobalTestSupply.com