Patient List on Programmer and the PDMS on a PC
•
When accessing the PDMS via a web browser (not via the tablet), you can define or edit
and save a detection set to the PDMS, and later download the detection set to the
programmer. Having saved a detection set to the programmer, you can program the
neurostimulator while the programmer is offline, as is required, for example, if an
operating room lacks network connectivity. Note that these instructions apply whether the
target programmer is a model PGM-300 (laptop) or model 5000 (RNS
•
S
TIMULATION
stimulation settings. You can adjust stimulation settings only using the tablet during an
interactive programming session. See Configure Responsive Therapy on page 70.
Figure 75 Stimulation screen through the patient list—review settings history only
®
RNS
System Programming Manual
screen: This screen enables you only to review the patient's history of
89
®
Tablet).