Import Neurostimulator Settings - Neuropace RNS 5000 User Programming Manual

Table of Contents

Advertisement

Tablet Instructions

Import Neurostimulator Settings

1. Interrogate the new neurostimulator, the one to which you will transfer settings.
2. Select the U
TILITIES
3. Select I
D
MPORT
available for import by device serial number and the date and time of last interrogation before the
settings were exported to the tablet. It also notifies you if the tablet has no settings that came
from a compatible neurostimulator (for example, if you exported from an RNS-320 for use on an
RNS-300M).
Figure 67 Import Device Settings dialog
Note: Device settings can be imported only once because they are deleted from the tablet once
you successfully import and select C
below.
4. Select the device settings—the selected settings will be highlighted—and then select I
dialog notifies you that it is importing device settings.
When complete, the tablet Home screen opens. It will include a red R
lower right.
Figure 68 Home screen with the red Review & Program tile at lower right
®
RNS
System Programming Manual
(toolbox) icon at upper right. The U
. The I
D
EVICE
MPORT
EVICE
ONFIRM
dialog opens (see figure above).
TILITIES
S
dialog appears. It shows the device settings
ETTINGS
P
from the R
ROGRAMMING
screen, as described
EVIEW
MPORT
& P
tile at
EVIEW
ROGRAM
82