Tablet Instructions
For troubleshooting information, refer to Abnormal Lead Impedance (greater than 3500 Ohms or
less than 250 Ohms) on page 92; and Insufficient Charge on page 95.
Significance of Battery Voltage
®
Normal RNS
Neurostimulator battery voltage measurements are between 2.38 and 3.20 Volts.
When the battery voltage measurement drops to 2.37 Volts or less, neurostimulator replacement is
recommended, and the tablet notifies you on screen by showing the elective replacement indicator
(ERI).
A
S
CTIVITY
CREEN
Select A
from the H
CTIVITY
Figure 11 Activity screen
Note: You must be logged in to the PDMS to access the A
N
P
L
EURO
ACE
OG
to the PDMS on page 15.) If you are not connected to the Internet, see Connect to the Internet
for PDMS Access on page 10.
The upper part of the A
events over time. Swipe left or right over the dates to view earlier or later activity. The lower part of the
screen shows (by default) a list of the N
over time, one row for each period of time between programming changes, which is called a programming
epoch. More view options are described below.
®
RNS
System Programming Manual
screen to open the A
OME
I
screen opens when you select A
N
screen is a histogram that shows the number of detection and stimulation
CTIVITY
EUROSTIM
screen.
CTIVITY
CTIVITY
, prompting you to log in. (See Log In
CTIVITY
S
—detection and stimulation settings—used
ETTINGS
screen. If you are not, the
23