Tablet Instructions
1. Select N
EUROSTIM
2. Select S
TIMULATION
screen opens. (Swipe left to see the right side.)
Figure 56 Stimulation screen, left side. Swipe left to see right side.
The screen has rows S
therapy is either configured or disabled. When one of the five therapies is disabled, all
successive therapies are disabled also—you cannot skip a therapy. Recall that detection
must be enabled before any therapy can be configured.
3. For each desired stimulation therapy and burst, starting with S
electrodes on each lead across which to deliver therapy, by making at least one electrode
positive [+] and another negative [–]. Note that the neurostimulator can is an optional
electrode.
In general, stimulation should be delivered to the leads and electrodes from which
electrographic patterns of interest are observed. For example, if electrographic activity of
interest is observed on all channels, then the stimulation pathway should be configured to
stimulate across all electrodes. However, if electrographic activity of interest is observed on
®
RNS
System Programming Manual
S
from the H
ETTINGS
OME
at upper center of the N
1 through S
TIM
TIM
screen.
S
screen. The S
EUROSTIM
ETTINGS
5 to program up to five stimulation therapies. Each
TIM
TIMULATION
1 B
1, select the
URST
71