Test stimulations should be delivered to each electrode individually (electrode to
neurostimulator) to determine whether there are acute side effects from stimulation.
Habituation to side effects resulting from stimulation has been observed; however it is not
advised to program settings that cause patient discomfort.
Avoid stimulation settings that cause an electrographic afterdischarge.
Caution:
Afterdischarge Activity
If evidence of afterdischarge activity resulting from stimulation is seen either on stored
ECoGs or during test stimulation delivery, stimulation parameters should be adjusted
to prevent such occurrence.
Enable Therapy: Review & Program
The new settings are not sent to the neurostimulator until you complete these steps.
1. Select R
EVIEW
changed settings. If you are satisfied with the new or changed settings, proceed to the next
step. If not satisfied, select C
Figure 64 Review screen after configuring therapy. Blue items have changed.
2. To activate the configured therapy, make sure S
center.
3. On the R
EVIEW
within approximately 1 inch of the neurostimulator and select C
satisfied, select C
®
RNS
System Programming Manual
& P
at upper right. The R
ROGRAM
at upper right to return to the previous screen.
ANCEL
screen, if you are satisfied with the new or changed settings, place the wand
to return to the previous screen.
ANCEL
screen opens, showing a list of all
EVIEW
is set to E
TIMULATION
NABLED
P
ONFIRM
78
at upper
. If not
ROGRAMMING