Neuropace RNS 5000 User Programming Manual page 69

Table of Contents

Advertisement

Tablet Instructions
2. You can select to share with:
a. A specific user, who could then access it on any programmer to which he is logged in.
Touch the S
(Available usernames would include those users who have been logged in to a
programmer while interrogating the neurostimulator associated with the patient to which
the detection set applies.)
b. All known programmers, which means any programmer that has interrogated the
neurostimulator associated with the patient to which the detection set applies
c. One specific programmer, which you must identify by its serial number. The serial
number is on the label affixed to the back of the programmer.
3. When you have selected a user or programmer(s) to share with, select the S
S
D
HARE
ETECTION
and will be sent to the indicated user and/or programmer(s) the next time they are synchronized
with the PDMS.
Figure 46 Example message reporting successful sharing
What Happens to a Shared Detection Set
Sharing a detection set using the foregoing method merely makes it available to the targeted user or
programmer when connected to the PDMS. Shared detection sets are not automatically downloaded
and saved locally to the programmer.
What happens is that the next time the targeted user or programmer connects to the PDMS, an icon
with two wavy red lines will appear on the N
N
I
dialog opens, where the alert icon is shown again with the text "A pending
EUROSTIM
NFO
Detection Set is available on PDMS. Click the icon to being loading it." When the user clicks this
icon, an overlay appears that describes the detection set: the name, the time it was created, and the
name of the person who shared it. They can A
P
D
ATTERN
ETECTION
set, which clears the alert and allows the user to continue with their session. Whether accepted or
skipped, the detection set will still be available in the list of detection sets in N
long as the programmer is connected to the PDMS. However, whether you accept or skip the
detection set, the programmer does not download the detection set to the programmer for offline use.
There is only one way to download a detection set to the programmer for offline use, described next.
®
RNS
System Programming Manual
field to select from a list of available usernames, if any.
HARE WITH USER
S
dialog. A message appears to tell you that the detection set was saved
ET
EUROSTIM
screen and automatically load the detection set, or they can S
I
tile. When they select the icon, a
NFO
the detection set, which will bring the user to the
CCEPT
65
button in the
HARE
the detection
KIP
S
—as
EUROSTIM
ETTINGS