Cal Mv Input - Fluke 724 Calibration Manual

Hide thumbs Also See for 724:
Table of Contents

Advertisement

724/725
Calibration Manual

Cal mV Input

Connect the UUT as shown in Figure 17.
Warning
30 V maximum to
SOURCE / MEASURE
mA +
3W
mA -
4W
From the Calibration Menu, type the number or letter for Cal mV Input. The PC
displays:
Set the Fluke 5520A to output 0.000 mV, let the output settle then press the space bar on
the PC. After a short while, the PC displays the following calibration constant and new
prompt:
Set the Fluke 5520A to output 90.000 mV, let the output settle, then press the space bar
on the PC. After a short while, the PC displays the following calibration constants and
new prompt:
Press the space bar to return to the Calibration Menu.
38
30V MAX ALL TERMINALS
MEASURE
V
V
TC
mA
Hz
RTD
LOOP
TC
COM
COM
UUT
Figure 17. mV Input Calibration Connections
Enter 0 mV
- press space bar to continue
Offset =
-714
Enter 90.00 mV
- press space bar to continue
diff = (Counts - Offset)
7513104
=
7512390
mV per count =
0.000012
-press space bar to continue
-
-714