Page 1
OPERATING MANUAL OPERATING MANUAL ba76199e03 08/2018 IDS WA-M IDS WA-S IDS WA-S Charger IDS WA System SYSTEM FOR THE WIRELESS OPERATION OF IDS SENSORS...
Page 4
7.2 Display of when a sensor is connected....... . . 20 7.3 The battery of the IDS WA-S adapter cannot be charged ....20 7.4 Status LED of the IDS WA-S Charger flashing red/green .
IDS sensor with plug head connector (variant W) and your IDS meter. Two adapters, one at the IDS meter (IDS WA-M) and one at the sensor (IDS WA- S), replace the sensor cable with an energy-saving Bluetooth LE radio connection.
2.2 Safe operation 2.2.1 Authorized use The authorized use of the IDS WA System is exclusively the running of a wireless connection beween IDS sensors with IDS meters. Only the operation and running of the analyzer according to the instructions and...
IDS WA System Commissioning mentioned in this operating manual. • The individual components of the IDS WA System must not be opened. Commissioning 3.1 Scope of delivery Depending on the order number, the scope of delivery consists of the whole IDS WA Kit, or one individual item out of the IDS WA Kit.
Charging contacts for the IDS WA-S adapter Retaining clip for the IDS WA-S adapter The IDS WA-S Charger is used both to charge and store the IDS WA-S adapters (e.g. overnight). The sensors may remain plugged to the IDS WA-S adapters while being charged.
WA-S Charger. The function keys of the adapter point upwards. 3. Close the retaining clip (5) of the IDS WA-S Charger until it clicks into place. The status LED (2) of the IDS WA-S Charger lights up red. The battery is being charged.
IDS meter. IDS WA-M 1. Turn the IDS WA-M adapter so that the notch on the plug of the IDS WA-S adapter is aligned to the guidance of the socket. 2. Plug the IDS WA-M adapter into an IDS sensor socket of the meter.
1. Open the locking device on the plug head sensor. 2. Align the guidance (1) of the IDS WA-S adapter to the notch (2) in the plug of the plug head sensor and insert the IDS WA-S adapter into the unlocked plug of the plug head sensor until is clicks into place.
Page 12
For information, the status indicator [RO] is displayed in the measured value display. ] BAT xx %: Current charging condition of the battery in the IDS WA-S adapter ] SIGNAL xx dBm: Current signal strength...
The measured value of the sensor is displayed. 3.7 Combining several IDS WA-S Charger modules To be able to charge several IDS WA-S adapters via one USB connection at the same time, you can combine up to 5 IDS WA-S Charger modules to form one charging station.
Page 14
Plug for the power supply of a further IDS WA-S Charger module via the socket (1) of the other module 1. Plug together up to 5 IDS WA-S Charger modules with the aid of the pin recep- tacle (3) and pin (4).
3.8 Mounting an IDS WA-S Charger module to a wall To store the IDS WA-S Charger and IDS WA-S adapter in a way that saves space and keeps them always ready for use, the IDS WA-S Charger is prepared to be...
Operation and measurement IDS WA System mounted to a wall. You can fix the IDS WA-S Charger to a wall in the following ways: Receptacle for hook Put the IDS WA-S Charger on a hook through the receptacle on the underside.
4.2 Saving the measured value 4.2.1 Functions of the IDS WA-S adapter The IDS WA-S adapter on the sensor has a function key with which simple functions can be executed on the sensor. Function key of the IDS WA-S adapter...
No data exchange, no connection Storing the IDS WA-S adapter For storing, put the IDS WA-S adapter with the sensor connected into a IDS WA-S Charger ready for operation. The battery of the IDS WA-S adapter is automatically recharged when the charge becomes less.
6.3 Disposal All components of the IDS WA System contain electronics. At the end of their operational lifetime, the components of the IDS WA System must be returned to the disposal or return system statutory in your country. If you have any questions, please contact your supplier.
– Fix the adapter with the retaining clip charging station – Battery exhaustively dis- – Carry out a reset of the IDS WA-S adapter in the charged charging station (IDS WA-S Charger): Press the function key of the adapter for at least 8...
IDS WA System Replacement parts and accessories 7.4 Status LED of the IDS WA-S Charger flashing red/green Cause Remedy – Charging error (e.g. bat- – Remove the IDS WA-S adapter from the charging tery is exhaustively dis- station charged) – Insert the IDS WA-S adapter in the charging sta- tion again –...
IDS WA System At the moment, there exist licenses of the BlueTooth LE radio module in use for Europe, USA, Canada, and other countries (list available from YSI on request.) Most important licenses: CE, FCC. All countries following these directives can use this product without hesitation.
IDS WA System Technical data Power supply Via the sensor connection socket of the meter 9.3 Adapter IDS WA-S Dimensions Approx. 83 x 20 x 20 mm Weight approx. 25 g Mechanical Type of IP 66 structure protection Test certificates...
Power supply USB connection and plug connections on the IDS WA- S Charger The IDS WA-S Charger requires a current of 180 mA to charge the battery of the IDS WA-S adapter USB 3.x connections of the PC meet this criterion...
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com...