DIAGNOSTICS
30JUL07
2
Do the following:
8F2919
Page
205 of 326
For open DOORS or
DRAWERS
CONNECTORS on the:
•
•
•
•
PDB
3
Install a new OPTICS AY. See the ADJUSTMENTS AND REPLACEMENTS,
and EXPOSURE TRANSPORT
Check:
DRE USB CABLE
PDB
ETB
SOP BOARD
If an existing problem:
Close all DOORS and DRAWERS.
Seat the CONNECTORS.
Install a new PDB. See the ADJUSTMENTS AND
REPLACEMENTS,
POWER DISTRIBUTION BOARD
(PDB).
AY.
Status Messages and Error Codes
OPTICS AY