14. Disinfection and Cleaning

Damage to ScanX Swift:
Do not spray disinfection or cleaning solutions on ScanX Swift.
Do not allow fluids to enter the unit.
Where the surfaces of the appliance show signs of contamination
or visible dirt, clean and disinfect the unit .
Switch ScanX Swift off by pressing the ON/OFF switch for 2
seconds .
Wait until the unit has run through the off cycle and has
completely shut down .
Clean the outside of ScanX Swift, dust cover and transfer box
using a damp cloth, and then use a disinfection wipe .
Unlock the inlet guide by pressing the locking catch and lift
the inlet guide upwards at the same time and remove .
Damage to ScanX Swift
No ScanX Swift parts should ever be disinfected using a
thermal disinfection unit or with the autoclave. The plastic
parts will be destroyed by the heat!
Internal parts can be cleaned using a damp cloth, and then
disinfected using a disinfection wipe .
Install the inlet guide .

15. Maintenance

15.1 Yearly Scheduled Maintenance.

Like all precision products, the ScanX Swift requires a certain
amount of care on a regularly scheduled basis . A well-organized
maintenance program aids dependable equipment operation
and reduces problems to a minimum . Yearly service carried out
by an Air Techniques trained and certified technician will ensure
continued uninterrupted operation . These routine checks help to
detect general overall wear, and replacement of parts can often be
made before a problem occurs .
15.2 Phosphor Storage Plates (PSPs).
PSP's are subject to "wear" on the black side during normal
handling and use . They can appear scratched, while the sensitive
blue side remains relatively "smooth" . This scratched look on the
black side has absolutely no effect on the quality of the image and
should be expected under normal conditions . If the phosphor side is
scratched make sure the plates are being handled properly and not
being dragged from ScanX Swift outlet pad area or other surfaces
that could cause scratching of the plate . Make sure to review the
Plates information provided by section 12 of this manual .
Page 22
Only use disinfection and cleaning agents specifically approved or
manufactured by Air Techniques. Other cleaning agents and disinfectants
could damage the ScanX Swift and its accessories.
Figure 14-1
Entry and Exit Cleaning
Figure 14-2
Inlet Guide Cleaning
Figure 14-3 .
Area to clean with Inlet Guide off