Operation

13.2 Scanner Start Up.

Switch on PC and monitor .
Switch on the ScanX Swift at the On/Off switch .
Start software .
Select the correct patient .
Set desired resolution .
Ready LED and status LED are both green .
Operational LEDs
Ready for operation
Idle - no communication
with software
Shut down in process
On
Off

13.3 Scanner Plate Processing Options.

Using the ScanX Swift, the user has the option to perform single plate processing or batch processing
depending on their needs . The difference between the two processing procedures is mainly how the
plates are handled . Refer to paragraphs 13 .4 and 13 .5 to perform single plate processing . Paragraphs
13 .6 and 13 .7 provide the batch processing procedures .
Page 18
After switching the unit on the ScanX Swift needs to go through initialization.
Ready for operation mode is shown by the green Ready LED being lit.
The green Status LED blinks while the ScanX Swift is being activated by
imaging software. As soon as the unit is ready for use, the Status LED
lights solid green.
When the On/Off switch is pressed for 2 seconds then the unit shuts down.
There may be a slight delay (green and blue LEDs blink alternately). The
unit will shut down completely. Do not move the scanner during this phase.
Operational LED
indicators
Blinking
Status LED
indicators
Status LEDs
Problem
Missing inlet guide
Plate currently being processed
Film feed slot ready; the next plate
can be inserted
Being activated by imaging software
On / Off
On
button
Blinking
Off
Air Techniques, Inc.