7.3 System Properties.

If unsure of the operating system version
installed, check that it meets the neces-
sary requirements by checking the System
Settings window as shown to the right .
The System Settings window can also be
opened from the Control Panel button . Just
press the Start button and select Control
Panel and then click the System icon .

8. Setup

8.1 Room for Set-up.

ScanX Swift should be set up in a closed, dry and well-ventilated room .
Room temperature should not fall below 50°F (+10°C) or exceed 95°F (+35°C) .
Do not install in a utility room, e .g . in a boiler room or in a wet room .
Set up in ordinary room light . Avoid direct sunlight and bright interior lighting (more than 800
lux) directly above the inlet guide .
There must be no source of interference in the room for set-up (e .g . strong magnetic field) .
Any such disturbances can interfere with the operation of the ScanX Swift .
8.2 Counter Top/Table Set-up.
ScanX Swift may be located almost anywhere in the office . Follow these guidelines for optimum performance:
Provide a stable, flat counter top large enough to hold the scanner .
Locate the computer within 6 feet (length of USB cord provided) .
Locate ScanX Swift no further than 6 feet from an AC outlet .
Make sure to leave the socket outlet easily accessible, so that the unit may be
unplugged if necessary . The power supply adapter is the Mains disconnect device .

8.3 Monitor.

Monitors designed for digital X-ray processes with a higher light intensity and broad contrast range
should be used . Strong room lighting, direct sunlight and reflections all serve to reduce the diagnostic
capabilities of the X-ray exposures .
Air Techniques, Inc.
Only suitably qualified personnel should set up, install
or operate ScanX Swift.
Damage through shock.
During operation and the switching off process (blue
and green operational LEDs blinking alternately) ensure
that the unit is not moved in any way whatsoever, as the
moving parts must be allowed to stop moving completely.
Danger of short circuit due to condensation.
ScanX Swift can only be put into operation once the
unit has warmed up to room temperature and is dry.
In order to avoid problems when scanning the image
data, ScanX Swift should be set up so that will not be
subject to knocks, vibration or similar.
Important Information