13.4 Single Plate Processing Scanner Preparation
Apply a barrier film to the ScanX Swift input guide making sure to
align the opening in the film center with the slot of the inlet guide .
See Figure 13-2 .

13.5 Feeding Single Plates.

Refer to Figure 13-3 and process a single plate as follows .
1 . Disinfect and Dry the Barrier Envelope (with plate still inside) .
2 . Hold the plate above the inlet guide with the printed side facing you and the notch of the envelope on
the bottom as shown .
3 . Tear the envelope lengthwise starting at the bottom notch while lowering the plate into the inlet guide .
4 . Press the plate all the way down until the unit takes over and moves the plate on its own .
5 . Remove and discard the envelope . The plate will drop into the bottom outlet pad when processing is
completed .
6 . View and save each scanned image using the user-supplied imaging software .
Air Techniques, Inc.
Contamination of the ScanX Swift
Use a new Barrier Film for each patient.
Loss of image data
Image data on the plate will be lost when exposed to light. Transfer the PSP
from the Barrier Envelope to the Plate Guide slot quickly. Always keep the
sensitive side away from any light source to minimize image loss.
Make sure that the user-supplied imaging software is set to scan, the
green control LED on the scanner is lit and the Swift Barrier Film is applied
View A
View C
Figure 13-3 . Single Plate Feeding and Scanning
Operation
Barrier Film
Figure 13-2 .
Applying barrier film
View B
View D