A&D ULTRACONNECT UB-1100BLE Instruction Manual page 9

Premium wireless wrist blood pressure monitor
Hide thumbs Also See for ULTRACONNECT UB-1100BLE:
Table of Contents

Advertisement

SYMBOLS
Symbols
Function/Meaning
The C.P.G. Indicator:
HIGH. The height of the
blood pressure monitor
is higher than your heart.
Appears while
measurement is in
progress.
The pulse is not detected
PUL
E
correctly.
/min
Unstable blood pressure
due to movement during
measurement. The systolic
Err 1
and diastolic values are
within 10 mmHg of each
other.
The pressure value
did not increase during
Err 3
inflation. The cuff is not
applied correctly.
Blood pressure monitor
Err 9
internal error
Data is transferring
Recommended Action
Place the blood pressure
monitor at the same level
as your heart.
Measurement is in
progress. Remain as
still as possible.
Apply the cuff correctly,
and take another
measurement.
Take another
measurement.
Remain still during
measurement.
Apply the cuff correctly,
and take another
measurement.
Press the reset button.
After reset, clock data
will disappear. If the error
still appears, contact
A&D Medical.
E-9