Ivc Bariatric Bed Specifications; Bed Components - Invacare IVC Bariatric Bed User Manual

Table of Contents

Advertisement

2.2 IVC Bariatric Bed Specifications

BED END HEIGHT
MINIMUM:
MAXIMUM:
DECK HEIGHT
MINIMUM:
(DOWN POSITION):
MAXIMUM (UP POSITION):
BED
OVERALL LENGTH:
WIDTH:
SPLIT SPRINGS TO ENDS
HEAD:
FOOT:
SHIPPING WEIGHT
WEIGHT CAPACITY
OCCUPANT:
RATED MAXIMUM:
(SAFE WORKING LOAD)
Height from floor to bed deck.
Measurements may vary ± 2 inches.
1123842-I-05
26 inches
34 inches
14 inches
22 inches
88 inches
42 inches
43 inches
45 inches
275 lbs
600 lbs
700 lbs (Including accessories,
mattress, occupant and any
other objects positioned on
the bed)
For houshold/residential or public occupancy use per UL
962.
Bed is NOT for hospital use and was not designed to meet
hospital standards.
Only use bed indoors.
WARNING!
Risk of Injury or Damage
The weight limit of the occupant of the bed CANNOT
exceed 600 pounds. To avoid injury and/or damage from
bed collapse:
– The TOTAL weight limit of the bed (including
accessories, mattress, occupant and any other objects
positioned on the bed) cannot exceed 700 pounds
(Safe Working Load).
WARNING!
Risk of Injury or Damage
Patients that weigh less than 351 pounds have an
increased risk of entrapment. To avoid injury or damage
from entrapment:
– Patients that weigh less than 351 pounds should not
use the IVC Bariatric Bed.

2.3 Bed Components

Head spring with one motor (BAR5000IVC)
Foot spring with two motors (BAR5490IVC)
Bed Ends (BAR5301IVC)
Overview
7

Advertisement

Table of Contents
loading