Other Settings - AOR SDU5600 Instruction Manual

Professional spectrum display unit for use with a companion receiver
Hide thumbs Also See for SDU5600:
Table of Contents

Advertisement

Using the numeric keypad, enter a new frequency followed by the [ENT] key ([KHz] key
or [MHz] key.)
Step Frequency (Ch. STEP)
Press the [Ch. STEP] key using the center Special Function Keys. The icon indicating
the step frequency will be reversed prompting the entry of new step frequency. Using the
numeric keypad, enter a new step frequency followed by the [ENT] key ([KHz] key or
[MHz] key.)
End (Stop) Frequency
Press the [Ch. END] key using the right Special Function Keys. The icon indicating the
end (stop)frequency will be reversed prompting the entry of a new end (stop) frequency.
Using the numeric keypad, enter a new end (stop)frequency followed by the [ENT] key
([KHz] key or [MHz] key.)
Receive Frequency (RX. FREQ)
After setting, the receiver will monitor the start frequency. To change the receive
frequency, move the marker using the rotary dial encoder and press the [MK.F] key.
[Memo: End (Stop) frequency limit:
It is calculated by the following formula.
(CH. START) + [(Ch. STEP) x 160] or (Ch. START) + 5 MHz

6-6-4 Other Settings

Reference Level (AMPLITUDE)
A 60 dB range is displayed on the scale which is divided into 10 dB increments.
The reference level can be selected between 0 dBm ~ -30 dBm in 10 dB steps
with 4 selectable steps.
Press the [AMPLITUDE] key located between the LCD display and the rotary dial
encoder. The icon indicating the current reference level will be reversed prompting a
new entry. Using either the numeric keypad or the rotary dial encoder, enter a new
reference level.
[Example: Reference level -20 dBm]
21