Error/Warning
Description
E2
During calibration,
the difference
between the
constants gets >
30%.
E3
During calibration,
the cell constant
-1
gets < 0.05 cm
.
E3
Slope out of range
(accepted values ±
58 mV).
E4
Buffer solution not
recognized.
Solution
• Insert the probe into the appropriate
standard and read again.
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
• Connect a new probe.
• Insert the probe into the appropriate
standard and read again.
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
Connect a new probe.
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
• Make sure that the buffer used
matches the buffer specified in
configuration.
• Replace the buffer solution.
• Make sure of the temperature
specification in configuration.
Error/Warning
Description
E4
Salinity cannot be
calculated TC=0
E5
Same buffer
solutions.
E6
The calibration
solutions have
different
temperatures.
ErA (only
Salinity is set to
MM156)
automatic and
measured value >
50 g/L.
E7
The datalog is
empty.
E8
The datalog is full.
E9
The data transfer
failed.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Solution
Modify TC
• Clean the probe and the membrane.
• Make sure that there are no air
bubbles in the membrane.
• Disconnect the probe and then
connect the probe again.
• Connect a different probe to verify if
problem is with probe or meter.
• Make sure that the buffer used
matches the buffer specified in
configuration.
• Replace the buffer solution.
• Make sure of the temperature
specification in configuration.
Make sure that the calibration solutions
have the same temperature.
Modify TC
Save data to the datalog.
Erase data from the datalog.
Make sure that the USB-RF device is
installed and that portCom is open.
English 17