Electrical Specifications - Amprobe CR50A User Manual

Capacitance, resistance meter
Hide thumbs Also See for CR50A:
Table of Contents

Advertisement

Available languages

Available languages

ELECTRICAL SPECIFICATIONS

(Accuracies at 23°C ± 5°C, <75% RH.)
RESISTANCE
Ranges
Accuracy
Resolution
OL Protection,
all Ranges
Open Circuit
Voltage
CAPACITANCE
Ranges
Accuracy
Test Frequency
20Ω, 200Ω, 2k-2MΩ, 20MΩ
±1.2% rdg.
±(0.5% rdg +3 dgt)
±(0.5%rdg + 1dgt)
±(3.0% rdg + 1dgt)
0.01 in 20Ω rg.
500VDC or AC RMS
0.3VDC; 3.0VDC on 20Ω
and 200Ω ranges
200pF, 2000pF to 2μF, 20μF,
200μF, 2mF, 20mF
±(0.5% rdg +1 dgt +0.5pF)
±(0.5% rdg +1 dgt)
±(0.5% rdg +1 dgt)
±(0.5% rdg +1 dgt)
±(1% rdg +1 dgt)
±(1.5% rdg +1 dgt)
820Hz, 820Hz, 82Hz, 8.2Hz,
8.2Hz, 8.2Hz
10