Check that the Ethernet cable is securely inserted into the spectrometer
and host computer. Check that the Ethernet LED is on.
•
For an isolated network, check that the Ethernet cable is a cross-over
Ethernet cable.
•
For an established network, check that the Ethernet cable is a standard
Ethernet cable.
•
Check that the IP Address is in the same range or subnet as the ASD
spectrometer. The same subnet means that the first three octets of the IP
address (xxx.xxx.xxx.___) match the spectrometer and the computer.
•
Do a ping test to make sure the spectrometer is responding.
»
Open up a command window by select Start->Run
»
Type cmd in the Run window.
»
Select OK to open the window.
For an Ethernet connection, type: ping 10.1.1.11
»
For a wireless connection, type: ping 10.1.1.77
»
Figure A-1 Run window with cmd.
74
Appendix A Troubleshooting
FieldSpec® 3 User Manual