Fore Optic Discussion - ASD FieldSpec 3 User Manual

Hide thumbs Also See for FieldSpec 3:
Table of Contents

Advertisement

www.asdi.com
4.3

Fore Optic Discussion

Field-of-View
ASD Document 600540 Rev. J
The instrument can be successfully used with a few broken fibers in each
range, although with a reduction in signal strength.
For severely damaged Fiberoptic cables, send the instrument back to ASD
Inc. in the carrying case for repairs.
The Fiberoptic cable provides the flexibility to adapt the instrument to a wide
range of applications. Field measurements can be performed by:
holding the pistol grip by hand, or
mounting the pistol grip and other foreoptic accessories to a tripod.
Reflected radiance and surface reflectance measurements are typically made
using the hand-held configuration. The pistol grip is available with both a
sighting scope and leveling device when required for more precise
orientation.
These accessories allow viewing the exact spot where the foreoptic is pointed
and orienting the foreoptic in precise, nadir-viewing, geometry. The majority
of irradiance measurements are performed with the irradiance receptor
mounted level on a tripod because of the need for precise geometric
orientation.
ASD's line of spectroradiometer is available with a wide selection of
foreoptics. These interchangeable foreoptics allow the flexibility to adapt the
FieldSpec spectroradiometer to a wide range of applications.
The small size of the FieldSpec's foreoptics allows positioning the foreoptics
at a greater distance from the surface under observation.
A field spectroradiometer with a larger field-of-view means that fewer
measurements are needed to approximate the spatial resolution of the imaging
sensor, because the pixel size of most imaging sensor systems is several
meters or more.
The small size of the pistol grip and foreoptics greatly reduce errors
associated with instrument self-shadowing. Even when the area viewed by the
foreoptic is outside the direct shadow of the spectroradiometer, the instrument
still blocks some of the illumination that would normally be striking the
surface under observation, either diffuse skylight or light scattered off
surrounding objects. Thus, the instrument as well as other objects —
including the user — should be placed as far as possible from the surface
under observation. This orientation requirement also applies to white
reference measurements.
33
Chapter 4 Fiberoptic Interface
FieldSpec® 3 User Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents