Temperature Influences During Measurements; Vapour Pressure Influence On Pf/Wc - UMS T5 User Manual

Pressure transducer tensiometer
Table of Contents

Advertisement

Useful notes

6.3 Temperature influences during measurements

If the sensor is not powered continuously the voltage should be
switched on 10 seconds before a measurement. In this case, the self
heating is negligible.
The correlation of water tension to water content is temperature
dependent. The influence is low at tensions of 0 to 10 kPa  0 ...
0,6 kPa/K, but high for tensions over 100 kPa:
 = Water tension R = Gas constant (8,31J/mol K)
M = Molecular weight
p
= Saturation vapour pressure at soil temperature
o
(from Scheffler/Straub, Grigull)

6.4 Vapour pressure influence on pF/WC

If the temperature of a soil with a constant water content rises from
20°C to 25°C the soil water tension is reduced for about 0,85 kPa
due to the increased vapour pressure which antagonizes the water
tension.
Temperature
in °C
Pressure
change per
Kelvin in [hPa]
 
R
M
p = Vapour pressure
4
10
16
0,6
0,9
1,2
T
 
ln
o
20
25
1,5
1,9
38/48
30
50
70
2,5
7,2
14