to measure such a signal level, but many loggers cannot do this.
In this case use a TV-batt power supply.
3.4.4
TV-batt Tensiometer power supply
The TV-batt power supply is specially designed for Tensiometers T3,
T4, T4e and T5. It offers a stabilized 10,6 V power supply, but with
supply minus = -5 V and supply plus = +5,6 V. Therefore the output
signal will have a logger specific signal level. The Tensiometer
signals are in a range of <1 V.
The TV-batt is directly supplied by battery or 12 V mains power.
3.4.5
Connection to a data logger
Some logger types can measure bridge circuits directly, other
loggers require certain measures as the Tensiometer signal minus
and the supply minus do not have the same ground.
3.4.6
Tensiometer loggers DL6-te or GP1-te
T5 and T5x can be connected directly and without further power
supply to the special Tensiometer loggers DL6-te or GP1-te.
The DL6-te is a stand-alone 6-channel logger with six 4-pin sockets.
All Tensiometer with 4-pin plug are connected to the DL6-te with
extension cables EC-4/...
The GP1-te is a stand-alone 2-channel logger with cable glands. All
Tensiometer with 4-pin plug are connected to the GP1-te with
connection cables CC-4/...