UMS T5 User Manual page 28

Pressure transducer tensiometer
Table of Contents

Advertisement

Service and maintenance
6. Pull up the syringe until the
spacers snap in.
The Tensiometer water now is
degassing.
Reassembling
1. Take of the tubes from the
shaft.
Taking off the syringe from the
sensor body needs care to avoid
damage of the pressure
transducer!
Leave both syringe assemblies
for at least 2 hours (the longer
the better).
The piston must not snap in
as
this
might
pressure transducer!
2. Hold the syringe and the
syringe piston securely. Bear in
mind there is still vacuum inside.
Press in the spacers and slowly
release the piston.
28/48
damage
the