Rae PGM62x6 User Manual page 48

Multirae series
Table of Contents

Advertisement

If you want to perform another snapshot benzene-specific measurement, press [Y/+] and
continue, using the same procedure as outlined above. Otherwise, press [N/-] to abort.
You see this screen:
8. If you are done performing benzene measurements or if you want to use the MultiRAE Benzene
for TVOC operation, either remove the RAE-Sep Tube Cartridge or press the button so that the
"B" is hidden and the "T" side is fully exposed (all the way out):
9. Press [N/-] to advance to the main reading screen.
Do not overtighten any portion of the sampling assembly.
Note: When the MultiRAE Benzene is used only for TVOC (total VOC) monitoring, there is no
need to use a RAE-Sep Tube Cartridge. If the cartridge is left on during VOC monitoring, make
sure it is set for TVOC measurement (the "T" is showing).
IMPORTANT!