Rae PGM62x6 User Manual page 129

Multirae series
Table of Contents

Advertisement

LEL Range, Resolution & Response Time
LEL
0-100%
LEL Correction Factors
Compound
Methane
Propane
Propene
n-Butane
Isobutylene
n-Pentane
n-Hexane
Cyclohexane
Benzene
Toluene
n-Heptane
n-Octane
Turpentine
Leaded Gasoline
Methanol
Ethanol
Isopropanol
Acetone
Methyl Ethyl Ketone
Ethyl Acetate
Carbon Monoxide
Hydrogen
Ammonia
* Response of the RAE Systems LEL sensor to a range of gases at the same LEL, expressed as percent of
methane response (=100). These figures are for guidance only and are rounded to the nearest 5%. For the
most accurate measurements, the instrument should be calibrated with the gas under investigation. See
RAE Systems Technical Note TN-156 for more details and other compounds.
Caution:
Refer to RAE Systems Technical Note TN-144 for LEL sensor poisoning.
1 %
15 sec
LEL Relative
Sensitivity*
100
62
67
50
67
45
43
40
45
38
42
34
34
48
67
59
38
45
38
45
75
91
125
LEL
CF
1.0
1.6
1.5
2.0
1.5
2.2
2.3
2.5
2.2
2.6
2.4
2.9
2.9
2.1
1.5
1.7
2.6
2.2
2.6
2.2
1.2
1.1
0.80
129

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pgm62x0