Calibration

9.4 Manual Calibration Procedure

To cancel calibration at any time, press
1.
From normal operation, press
to select Calibration.
2.
In the Calibration menu, scroll down and select Calibration.
3.
The detector zeroes the sensor before the calibration.
28
.
to enter Configuration Mode. Press
4.
When the zero calibration is complete, attach the calibration cap and apply
the calibration gas at a flow rate of 0.25-0.50 L/min.
Gas concentration 25ppm
If Cal IR Lock is enabled, the following screen displays to indicate calibration
can only be performed using an IR device (IntelliDoX or IR Link Adapter).
5.
Once the detector has detected a sufficient amount of gas, the detector
begins calibration of the sensor.
Apply gas now
waiting... 300
ppm - H2S
25ppm
Calibrating