Set The Sound Options; Security Options; Turn Security Options On; View Meter Information - Hach 2100Qis User Manual

Hide thumbs Also See for 2100Qis:
Table of Contents

Advertisement

Set the sound options

The meter can make an audible sound when a key is pushed, when a
reading is complete or when the calibration reminder is due.
1. Push SETTINGS and select Sounds.
2. Select which events will produce an audible sound. Multiple items
can be selected.
Option
Description
Key Press
The meter will make an audible sound whenever a key
is pushed.
Reading complete The meter will make an audible sound whenever a
reading is completed.
Reminders
The meter will make an audible sound when a
calibration is due.

Security options

The Security Options menu is used to protect the meter setup.
The Setup Date and Time, Delete Data Log, Restoring Factory Defaults
and Restore Factory Cal screens are not accessible without a password.
Store the password in a safe and accessible place. If the specified
password is forgotten and Security Options is turned on, the operator is
locked out of the restricted menus. Contact technical support if the
password is lost.

Turn security options on

The security options and the set password options are used together to
prevent access to restricted menus.
1. Push the SETTINGS key and select Security Options.
2. Select Edit Password and use the UP and DOWN keys to set a
password.
3. Select Security On to enable the password setting. The requirement
for the password entry is controlled by setting Security Options on or
off.
Note: Set the Security to Off to disable the password setting.
4. Push the ON/OFF key to turn off and on the meter to activate the
password settings.

View meter information

The instrument information menu shows specific information such as the
meter name, model number, software version, serial number and
available Operator and Sample IDs. 10 Operator IDs and 100 Sample
IDs are available.
1. Push the SETTINGS key and select Meter Information.

Calibration

The portable turbidimeter is calibrated with Formazin Primary Standards
at the factory. The meter should be calibrated upon receipt for best
results. The manufacturer recommends calibration with a primary
®
standard such as StablCal
Stabilized Standards or with formazin
standards every three months.
Note: Set Cal Reminder Repeat in the Calibration Options menu for periodical
calibration. Verify the calibration once a week.

Calibration options

The calibration options contain Calibration History, Calibration Curves,
Cal Reminder Repeat and Restore Factory Calibration.
1. Push the CALIBRATION key and then the UPand DOWN key.
Option
Description
Calibration
The calibration history shows a list of the times when the
History
meter was calibrated. Select a date and time to view a
summary of the calibration data.
English 13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2100q

Table of Contents